m+Medelevator.RU
← Поиск публикаций
Contact dermatitis · 2017

Allergic contact dermatitis caused by isobornyl acrylate in Freestyle® Libre, a newly introduced glucose sensor.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Herman Anne, Aerts Olivier, Baeck Marie, Bruze Magnus, De Block Christophe, Goossens An, Hamnerius Nils, Huygens Sara, Maiter Dominique, Tennstedt Dominique, Vandeleene Bernard, Mowitz Martin

Аннотация

BACKGROUND: Glucose sensors, such as FreeStyle® Libre, are innovative medical devices developed for diabetes patients as a replacement for classic glucose meters, ensuring continuous glucose monitoring without the disadvantage of regular skin finger pricks.

OBJECTIVES: To report several cases of allergic contact dermatitis caused by FreeStyle® Libre, and to report on isobornyl acrylate as a culprit allergen.

PATIENTS AND METHODS: Fifteen patients presented with allergic contact dermatitis caused by FreeStyle® Libre. All but 1 were patch tested with a baseline series, and with pieces and/or ultrasonic bath extracts of (the adhesive part of) the glucose sensor. Isobornyl acrylate was patch tested, in various concentrations and vehicles, in 13 patients. Gas chromatography-mass spectrometry (GC-MS) of the sensors was performed.

RESULTS: All patients reacted to the adhesive part of the sensor, and 12 patients were shown to be sensitized to isobornyl acrylate. Simultaneous reactions to other allergens were rarely observed. GC-MS showed the presence of isobornyl acrylate in the sensors.

CONCLUSIONS: Cases of allergic contact dermatitis caused by FreeStyle® Libre are increasingly being observed, and isobornyl acrylate is a relevant culprit allergen. Cross-reactivity to other acrylates was infrequently observed, but other, hitherto unidentified, contact allergens may still be present in the device.