BACKGROUND
Electrosurgical devices are widely used in penile surgery for hemostasis and tissue dissection. However, the neurovascular risks associated with monopolar electrocautery, particularly in the context of pelvic innervation and vascular supply, remain unclear. This study investigates the histopathological effects of monopolar versus bipolar electrocautery on pudendal ganglia and prostatic arteries in an experimental model.
MATERIALS AND
METHODS
Seventeen male New Zealand rabbits were randomly assigned to three groups: Group I (control, surgical dissection = 5), Group II (bipolar electrocautery = 5), and Group III (monopolar electrocautery = 7). A standardized preputial incision was performed. After a 3-week follow-up, pudendal ganglia and prostatic arteries were excised and analyzed. Degenerated neuron density (DND) and vasospasm index (VSI) were evaluated stereologically. The relatively small sample size (n = 17) represents a methodological limitation and may affect the generalizability of the findings.
RESULTS
Quantitative analyses revealed significantly higher DND and VSI in the monopolar group (DND: 198 ± 37 neurons/mm³; VSI: 2.234 ± 0.987), compared to the bipolar (DND: 31 ± 6; VSI: 1.197 ± 0.120) and control groups (DND: 9 ± 3; VSI: 1.034 ± 0.036), with all comparisons showing statistical significance (p < 0.001). These findings indicate progressive neurovascular injury, most pronounced with monopolar electrocautery.
CONCLUSIONS
Monopolar electrocautery during penile surgery induces marked neurovascular injury, potentially impairing pelvic nerve function and vascular regulation. Bipolar cautery is associated with a safer histological profile and should be preferred. Future research should include molecular, functional, and long-term outcome assessments.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.