Influence of the glenoid baseplate position on the direction and length of the superior and inferior locking screws.
Аннотация
Introduction
Superior screw insertion in reverse shoulder arthroplasty (RSA) carries the potential risk of suprascapular injury. The purpose of this study was to evaluate how the baseplate position affects the superior screw position and length in RSA.
Methods
Three-dimensional (3D) computer simulation models of RSA were established using computed tomography data of baseplates with superior and inferior screws and 3D scapular models from 10 fresh cadavers. Superior screw position, the distance from the superior screw hole to the suprascapular notch, and the screw lengths were measured and compared among various baseplate positions with two inferior tilts (0 and 10 degrees) and three rotational patterns (11-5, 12-6, and 1-7 o'clock in the right shoulder).
Results
For the 1-7 o'clock / inferior tilt 0 degrees baseplate, the superior screw located anterior to the SS notch in all shoulders, the distance to the SS notch was the longest (12.8 mm), and the inferior screw length was the shortest (23.1 mm).
Conclusion
Although there is a concern of a short inferior screw length, initial fixation using a baseplate with 1-7 o'clock rotation and an inferior tilt of 0 degrees appears preferable for SS nerve injury prevention during superior screw insertion. J. Med. Invest. 69 : 185-190, August, 2022.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.