Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid-liquid extraction method for the ABSCIEX API-5000 mass spectrometry system.

Van Der Gugten Jessica Grace, Dubland Joshua, Liu Hua-Fen, Wang Alex, Joseph Christine, Holmes Daniel T

СтатьяПеревод в очереди

Аннотация

AIMS: Accurate serum aldosterone determination is critical to the screening and diagnosis of primary aldosteronism, the localisation of aldosterone producing tumours, and the investigation of other disorders of the renin-angiotensin system. Mass spectrometry offers a means to overcome problems with method-dependent bias between competitive immunoassays for aldosterone. The authors have developed a simple, sensitive and precise liquid-liquid extraction aldosterone method for the ABSCIEX API-5000 liquid chromatography and tandem mass spectrometry (LC-MS/MS) system.

METHODS

Using d7-aldosterone internal standard, 500 μl of sample is extracted with 2500 μl of methyl tertbutyl ether followed by dry-down, reconstitution and LC-MS/MS analysis in ESI negative mode. Method validation was undertaken using standard approaches and comparison made against a commercial radioimmunoassay. Accuracy was assessed using EQA material with assigned aldosterone concentrations.

RESULTS

The assay was linear up to 3420 pmol/l (LOQ=50 pmol/l, LOD<22 pmol/l). Total CVs were ≤5% for concentrations ≥120 pmol/l and 10% at the LOQ. Mean accuracy was 98.5% against GCMS assigned material.

CONCLUSION

The authors present a precise, sensitive and simple aldosterone method suitable for routine clinical use that requires no solid phase extraction or specialised ion sources.

Для цитирования: Van Der Gugten Jessica Grace, Dubland Joshua, Liu Hua-Fen, Wang Alex, Joseph Christine, Holmes Daniel T. Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid-liquid extraction method for the ABSCIEX API-5000 mass spectrometry system.. Journal of clinical pathology. 2012. DOI: 10.1136/jclinpath-2011-200564.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.