DOI 10.1016/j.jcin.2024.05.039

Медицинская библиотека · JACC. Cardiovascular interventions · 2024

Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses: Implications for Valve Hemodynamic and Clinical Outcomes.

Maznyczka Annette, Heg Dierik, Tomii Daijiro, Nakase Masaaki, Baekke Pernille Steen, Lanz Jonas, Stortecky Stefan, Reineke David, Windecker Stephan, Pilgrim Thomas

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Asymmetrical expansion of transcatheter heart valves (THVs), manifesting as stent frame deformation, is an occasional fluoroscopic finding in transcatheter aortic valve replacement (TAVR).

OBJECTIVES

The aim of this study was to investigate the impact of asymmetrical expansion of balloon-expandable THVs on hemodynamic valve performance and clinical outcomes.

METHODS

In a prospective registry, TAVR asymmetry index was measured using freeze-frame fluoroscopic images and was defined as the ratio of THV heights: [(longer height/shorter height) - 1] × 100. THV hemodynamic performance was measured using echocardiography before hospital discharge. Impaired hemodynamic valve performance was defined as a mean residual THV gradient ≥20 mm Hg and/or moderate or greater paravalvular regurgitation.

RESULTS

Among 1,216 patients undergoing transfemoral TAVR for native severe aortic valve stenosis with contemporary balloon-expandable THVs between February 2014 and June 2022, asymmetry index was an excellent predictor of impaired hemodynamic valve performance (area under the receiver-operating characteristic curve: 0.88; 95% CI: 0.84-0.92; P < 0.001). The optimal asymmetry index threshold for predicting impaired THV performance was >5.5% (sensitivity 77%, specificity 86%) and occurred in 17% of patients. Higher asymmetry index, as a continuous variable, was associated with impaired hemodynamic valve performance independent of total aortic valve calcium, bicuspid anatomy, balloon-expandable valve prosthesis type or size, and Society of Thoracic Surgeons Predicted Risk of Mortality score (OR: 1.37; 95% CI: 1.29-1.46; P < 0.001). High TAVR asymmetry index was not associated with all-cause mortality during a median of 376 days of follow-up (HR: 1.00; 95% CI: 0.71-1.41; P = 0.989).

CONCLUSIONS

Asymmetrical expansion of balloon-expandable THVs was associated with impaired hemodynamic valve performance but not with clinical outcomes.

Для цитирования: Maznyczka Annette, Heg Dierik, Tomii Daijiro, Nakase Masaaki, Baekke Pernille Steen, Lanz Jonas, Stortecky Stefan, Reineke David, Windecker Stephan, Pilgrim Thomas. Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses: Implications for Valve Hemodynamic and Clinical Outcomes.. JACC. Cardiovascular interventions. 2024. DOI: 10.1016/j.jcin.2024.05.039.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.