Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial.
Аннотация
PURPOSE: To evaluate the safety and effectiveness of the Pipeline Embolization Device (PED; ev3/Covidien, Irvine, Calif) in the treatment of complex intracranial aneurysms.
MATERIALS AND
METHODS
The Pipeline for Uncoilable or Failed Aneurysms is a multicenter, prospective, interventional, single-arm trial of PED for the treatment of uncoilable or failed aneurysms of the internal carotid artery. Institutional review board approval of the HIPAA-compliant study protocol was obtained from each center. After providing informed consent, 108 patients with recently unruptured large and giant wide-necked aneurysms were enrolled in the study. The primary effectiveness endpoint was angiographic evaluation that demonstrated complete aneurysm occlusion and absence of major stenosis at 180 days. The primary safety endpoint was occurrence of major ipsilateral stroke or neurologic death at 180 days.
RESULTS
PED placement was technically successful in 107 of 108 patients (99.1%). Mean aneurysm size was 18.2 mm; 22 aneurysms (20.4%) were giant (>25 mm). Of the 106 aneurysms, 78 met the study's primary effectiveness endpoint (73.6%; 95% posterior probability interval: 64.4%-81.0%). Six of the 107 patients in the safety cohort experienced a major ipsilateral stroke or neurologic death (5.6%; 95% posterior probability interval: 2.6%-11.7%).
CONCLUSION
PED offers a reasonably safe and effective treatment of large or giant intracranial internal carotid artery aneurysms, demonstrated by high rates of complete aneurysm occlusion and low rates of adverse neurologic events; even in aneurysms failing previous alternative treatments.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.