Combined diagnostic accuracy of two artificial intelligence systems for glaucoma diagnosis using color fundus photography.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
da Silva Filho Gilvan Vilarinho, Caiado Gustavo Coelho, Yamamoto Felipe Zocatelli, Teixeira Sergio Henrique, Prata Tiago Dos Santos, Gracitelli Carolina Pelegrini Barbosa, Paranhos Augusto
Аннотация
PURPOSE: To evaluate the diagnostic accuracy of two commercially available artificial intelligence (AI) systems based on color fundus photography (CFP), the Laguna ONhE and VUNO Med-Fundus AI, for glaucoma detection, both independently and in combination.
METHODS: This retrospective cross-sectional study included 370 eyes from 193 patients (248 eyes with primary open-angle glaucoma and 122 healthy eyes). All eyes underwent structural evaluation with swept-source optical coherence tomography (Triton, Topcon) and visual field testing (Octopus 900, Haag-Streit AG). Fundus photographs were analyzed using Laguna ONhE and VUNO Med-Fundus AI systems. Diagnostic accuracy was evaluated.
RESULTS: Both AI systems demonstrated high diagnostic accuracy. Laguna achieved an AUC of 0.879 using Glaucoma Discriminant Function (GDF), and VUNO showed an AUC of 0.857. When combined, GDF + VUNO achieved an AUC of 0.903. The Global Mean Deviation (GMD) reached the highest diagnostic accuracy (AUC = 0.916), which was not significantly different from GDF + VUNO (p = 0.146).
CONCLUSIONS: Laguna ONhE and VUNO Med-Fundus AI had high diagnostic accuracy in detecting glaucoma using only CFP. Their combined use improved further, achieving accuracy comparable to the GMD. This represents a practical approach for glaucoma screening, particularly in settings without access to OCT or automated perimetry.
METHODS: This retrospective cross-sectional study included 370 eyes from 193 patients (248 eyes with primary open-angle glaucoma and 122 healthy eyes). All eyes underwent structural evaluation with swept-source optical coherence tomography (Triton, Topcon) and visual field testing (Octopus 900, Haag-Streit AG). Fundus photographs were analyzed using Laguna ONhE and VUNO Med-Fundus AI systems. Diagnostic accuracy was evaluated.
RESULTS: Both AI systems demonstrated high diagnostic accuracy. Laguna achieved an AUC of 0.879 using Glaucoma Discriminant Function (GDF), and VUNO showed an AUC of 0.857. When combined, GDF + VUNO achieved an AUC of 0.903. The Global Mean Deviation (GMD) reached the highest diagnostic accuracy (AUC = 0.916), which was not significantly different from GDF + VUNO (p = 0.146).
CONCLUSIONS: Laguna ONhE and VUNO Med-Fundus AI had high diagnostic accuracy in detecting glaucoma using only CFP. Their combined use improved further, achieving accuracy comparable to the GMD. This represents a practical approach for glaucoma screening, particularly in settings without access to OCT or automated perimetry.