DOI 10.1016/j.ijid.2018.06.005
Immunochromatographic test for rapid serological diagnosis of human angiostrongyliasis.
Аннотация
OBJECTIVES
The serological diagnosis of human infection with Angiostrongylus cantonensis remains problematic because there are no commercially available validated tests. Most laboratories use domestically prepared tests such as the enzyme-linked immunosorbent assay (ELISA) or immunoblotting. Since laboratory facilities are not always available in endemic areas, we developed and assessed a rapid lateral flow immunochromatographic assay (AcQuickDx Test) to detect anti-A. cantonensis antibodies in human serum.
METHODS
The test device was assembled with purified 31-kDa glycoprotein as diagnostic antigen and with gold-labelled anti-human immunoglublin-G as the detector reagent. A total of 97 serum samples were tested - 19 samples from clinically diagnosed patients with detectable A. cantonensis-specific antibody in immunoblotting; 43 samples from patients with other parasitic diseases, i.e. gnathostomiasis (n=13), toxocariasis (n=2), trichinellosis (n=2), hookworm infection (n=4), filariasis (n=5), cysticercosis (n=9), paragonimiasis (n=2), opisthorchiasis (n=3), and malaria (n=3); and 35 samples from normal healthy subjects.
RESULTS
The sensitivity, specificity, positive predictive value and negative predictive value of AcQuickDx Test to detect anti-A. cantonensis specific antibodies in serologically confirmed angiostrongyliasis cases, were 100%, 98.72%, 95% and 100%, respectively. Positive AcQuickDx was observed in 1 of 4 cases with hookworm infections. No positive AcQuickDx was observed in cases with other parasitic diseases, and the individual healthy subjects.
CONCLUSIONS
AcQuickDx Test is rapid, highly sensitive and specific, and easy to perform without additional equipment or ancillary supplies. It yields results that are interpreted visually, and possesses a long shelf-life at room temperature. Thus, it can be applied as an additional test for clinical diagnostic support of angiostrongyliasis either in conventional laboratories or for remote areas where laboratory infrastructure is not available.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.