DOI 10.3390/bioengineering13030296

Медицинская библиотека · Bioengineering (Basel, Switzerland) · 2026

Interdevice Agreement of Keratometry, Astigmatism Vectors, and Ocular Biometry in Cataract Candidates: SS-OCT (Argos) vs. OLCI (Aladdin) vs. Scheimpflug-Placido (Sirius).

Al Barri Leila, Yasar Ionela-Iasmina, Mercea Nadina, Tudor Anca, Stanca Horia T, Roșca Cosmin, Munteanu Mihnea

СтатьяПеревод в очереди

Аннотация

Background and

Objectives

Accurate anterior segment measurements are central to intraocular lens (IOL) power calculation and toric planning, yet different optical platforms may yield non-interchangeable values. This study compared keratometry, astigmatism metrics, and ocular biometry obtained with a swept-source OCT biometer (Argos), an optical low-coherence interferometry biometer (Aladdin), and a combined Scheimpflug-Placido topographer (Schwind Sirius).

Methods

This is a retrospective observational study (January 2022-June 2024) including eyes undergoing uncomplicated cataract surgery. All eyes were measured in a single session by one examiner. Outcomes included K1, K2, cylinder, astigmatism axis (degrees; device-reported corneal cylinder axis, labeled "Powerful Angle" in the Sirius export), vector components (J0 and J45), and-where available-lens thickness (LT), axial length (AL), anterior chamber depth (ACD), white-to-white (WTW) distance, and central corneal thickness (CCT). Friedman tests assessed 3-device differences, and pairwise comparisons were evaluated using Wilcoxon signed-rank tests (paired data).

Results

A total of 170 eyes (102 patients) were analyzed (mean age: 69.12 ± 10.26 years). Significant inter-device differences were detected for K1 (Argos: 43.45 ± 1.64 D; Aladdin: 43.41 ± 1.70 D; overall: p < 0.001; Argos vs. Aladdin: p = 0.019), K2 (Argos: 44.45 ± 1.67 D; Aladdin: 44.34 ± 1.71 D; overall and pairwise: p < 0.001), and cylinder (Argos: -0.83 ± 0.74 D, Aladdin: -0.77 ± 0.76 D; Sirius: -0.68 ± 0.75 D; overall: p < 0.001). "Powerful Angle" differed across devices (p = 0.003) but not between Argos and Aladdin (p = 0.512). J0 (p = 0.277) and J45 (p = 0.084) did not differ significantly. Argos reported higher ACD (3.19 ± 0.42 vs. 3.13 ± 0.41 mm, p < 0.001) and WTW (11.95 ± 0.42 vs. 11.65 ± 0.39 mm, p < 0.001) values than Aladdin. CCT was similar between Aladdin and Sirius (540.27 ± 33.44 vs. 540.47 ± 33.78 µm, p = 0.169).

Conclusions

Several keratometric and biometric parameters differed significantly by device, indicating limited interchangeability-particularly relevant for toric and premium IOL planning-while vector astigmatism components and CCT showed better agreement.

Для цитирования: Al Barri Leila, Yasar Ionela-Iasmina, Mercea Nadina, Tudor Anca, Stanca Horia T, Roșca Cosmin, Munteanu Mihnea. Interdevice Agreement of Keratometry, Astigmatism Vectors, and Ocular Biometry in Cataract Candidates: SS-OCT (Argos) vs. OLCI (Aladdin) vs. Scheimpflug-Placido (Sirius).. Bioengineering (Basel, Switzerland). 2026. DOI: 10.3390/bioengineering13030296.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.