BACKGROUND

Recently, automated alerts have been used to identify patients with respiratory failure based on set criteria, which can be gleaned from the electronic medical record (EMR). Such an approach may also be useful for identifying patients with severe adult respiratory distress syndrome (ARDS) who may benefit from extracorporeal membrane oxygenation (ECMO). Inhaled nitric oxide (iNO) is a common rescue therapy for severe ARDS which can be easily tracked in the EMR, and some patients started on iNO may have indications for initiating ECMO. This case series summarizes our experience with using automated electronic alerts for ECMO team activation focused particularly on an alert triggered by the initiation of iNO.

METHODS

After a brief trial evaluation, our Smart Alert system generated an automated page and e-mail alert to ECMO team members whenever a nonzero value for iNO appeared in the respiratory care section of our EMR. If iNO was initiated for severe respiratory failure, a detailed evaluation by the ECMO team determined if ECMO was indicated. For those patients managed with ECMO, we tabulated baseline characteristics, indication for ECMO, and outcomes.

RESULTS

From September 2012 to July 2013, 45 iNO alerts were generated on 42 unique patients. Six patients (14%) met criteria for ECMO. Of these, four were identified exclusively by the iNO alert. At the time of the alert, the median PaO₂-to-FIO₂ ratio was 64 mm Hg (range, 55-107 mm Hg), the median age-adjusted oxygenation index was 73 (range, 51-96), and the median Murray score was 3.4 (range, 3-3.75), indicating severe respiratory failure. Median time from iNO alert to ECMO initiation was 81 hours (range, -2-292 hours). Survival to hospital discharge was 83% in those managed with ECMO.

CONCLUSION

Automated alerts may be useful for identifying patients with severe ARDS who may be ECMO candidates.

LEVEL OF EVIDENCE: Diagnostic test, level V.

Для цитирования: Belenkiy Slava M, Batchinsky Andriy I, Park Timothy S, Luellen David E, Serio-Melvin Maria L, Cancio Leopoldo C, Pamplin Jeremy C, Chung Kevin K, Salinas Josè, Cannon Jeremy W. Automated inhaled nitric oxide alerts for adult extracorporeal membrane oxygenation patient identification.. The journal of trauma and acute care surgery. 2014. DOI: 10.1097/ta.0000000000000343.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.