American journal of ophthalmology · 2006
Pigment release and secondary glaucoma after implantation of single-piece acrylic intraocular lenses in the ciliary sulcus.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Uy Harvey Siy, Chan Pik Sha T
Аннотация
PURPOSE: To report the association of sulcus-fixated, single-piece hydrophobic acrylic intraocular lenses (HAIOL) with pigment release and secondary glaucoma.
DESIGN: Interventional case series.
METHODS: HAIOL was implanted in the ciliary sulcus of 20 eyes that developed posterior capsule rupture during phacoemulsification. We analyzed postoperative best-corrected visual acuity, manifest refraction, frequency of intraocular lens malpositioning, and postoperative complications.
RESULTS: Postoperative best-corrected visual acuity was 20/40 or better in all eyes. The mean postoperative sphere was -0.5 +/- 0.7 diopters (range +1.25 to -2.00); the mean postoperative cylinder was -1.2 +/- 0.8 diopters (range 0 to -2.50). None of the HAIOLs was decentered or dislocated, and none required repositioning. Seven eyes (35%) developed pigment release; three eyes (15%) developed secondary glaucoma.
CONCLUSIONS: Implantation of single-piece HAIOLs in the ciliary sulcus is associated with pigment release and secondary glaucoma. Alternative methods of aphakic correction should be considered.
DESIGN: Interventional case series.
METHODS: HAIOL was implanted in the ciliary sulcus of 20 eyes that developed posterior capsule rupture during phacoemulsification. We analyzed postoperative best-corrected visual acuity, manifest refraction, frequency of intraocular lens malpositioning, and postoperative complications.
RESULTS: Postoperative best-corrected visual acuity was 20/40 or better in all eyes. The mean postoperative sphere was -0.5 +/- 0.7 diopters (range +1.25 to -2.00); the mean postoperative cylinder was -1.2 +/- 0.8 diopters (range 0 to -2.50). None of the HAIOLs was decentered or dislocated, and none required repositioning. Seven eyes (35%) developed pigment release; three eyes (15%) developed secondary glaucoma.
CONCLUSIONS: Implantation of single-piece HAIOLs in the ciliary sulcus is associated with pigment release and secondary glaucoma. Alternative methods of aphakic correction should be considered.