DOI 10.1016/j.jaccas.2026.108495
Descending Aorta Perforation During Transseptal Puncture for Left Atrial Appendage Closure.
Аннотация
BACKGROUND
Aortic root injury during transseptal puncture (TSP) is a rare but well-recognized complication. However, injury to the descending aorta has not previously been described in this context.
CASE SUMMARY: A 57-year-old woman underwent left atrial appendage closure. During TSP using the VersaCross Connect radiofrequency-powered system and a 15-F steerable sheath, an unrecognized perforation of the descending aorta occurred. Despite repeated transthoracic echocardiograms showing no pericardial effusion and initial hemodynamic stability, she subsequently developed a mediastinal hematoma and bilateral pleural effusion. Management required an emergency percutaneous thoracic endovascular aortic repair, followed by surgical evacuation of the hematoma.
DISCUSSION: Excessive tenting during radiofrequency-powered TSP, related to interatrial septum aneurysm, resulted in injury to the descending aorta with life-threatening consequences.
TAKE-HOME MESSAGE: Radiofrequency-powered systems for TSP have enhanced tissue penetration capability and should therefore be used with caution, particularly in the presence of an interatrial septum aneurysm and atypical echocardiographic windows.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.