Refractive outcomes and accuracy of IOL power calculation with the SRK/T formula for sutured, scleral-fixated Akreos AO60 intraocular lenses.

Ohr Matthew P, Wisely C Ellis

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Scleral fixation of intraocular lenses has become a popular procedure for treating aphakia in the absence of capsular support. However, the lens formulas used to predict refractive outcomes were designed for in-the-bag lens placement. This study evaluates the accuracy of the SRK/T formula in predicting a target postoperative refraction when suturing a scleral-fixated intraocular lens (IOL) implant 3 mm posterior to the limbus.

METHODS

This is a retrospective, case series including 20 eyes of 20 patients who underwent scleral fixation of Akreos AO60 IOLs (Bausch & Lomb, Rochester, NY) by a single surgeon at the OSU Wexner Medical Center. Preoperative measurements were performed with optical biometry, and IOL power was calculated with the SRK/T formula. Following surgery, the actual refractive spherical equivalent (SE) was performed and compared with the preoperative prediction. Prediction error (PE), defined as the deviation of actual postoperative SE refraction in diopters (D) from preoperative predicted SE refraction, was the primary outcome measure.

RESULTS

The mean attempted (predicted) SE was - 1.12 D (± 0.87). Mean achieved SE was - 0.96 D (± 1.04). Mean PE (actual postoperative SE versus predicted preoperative SE) was 0.16 D (± 0.69). A total of 9 eyes (45%) were within ± 0.5 D of the predicted SE, 16 eyes (80%) were within ± 1.0 D, and all 20 eyes (100%) were within ± 1.5 D.

CONCLUSION

IOL power calculation using the SRK/T formula with optical biometry demonstrates reliable postoperative refractive outcomes in patients undergoing scleral fixation of an IOL (Akreos AO60). Further studies are needed to refine the predictive value of the SRK/T and other formulas for application in scleral fixation of IOLs.

Для цитирования: Ohr Matthew P, Wisely C Ellis. Refractive outcomes and accuracy of IOL power calculation with the SRK/T formula for sutured, scleral-fixated Akreos AO60 intraocular lenses.. Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie. 2020. DOI: 10.1007/s00417-020-04721-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.