Evaluation of a canine hemoglobin A1c point-of-care assay: analytical validation, clinical utility, and establishing a population-based reference interval.

СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Adam Denzel, Noto Nicholas, Rosser Michael F, Connolly Sarah, Gal Arnon

Аннотация

BACKGROUND: Hemoglobin A1c (HbA1c) assay validation is essential for establishing analytical reliability, evaluating discrimination between diabetic and non-diabetic (NDM) dogs, and establishing reference intervals.

HYPOTHESIS/OBJECTIVES: Validate a canine HbA1c point-of-care analyzer (Vet Chroma™), evaluate its diagnostic performance, and establish a reference interval in healthy dogs.

ANIMALS: Fifty-one diabetic, 44 NDM controls, and 42 healthy client-owned dogs.

METHODS: Prospective validation of precision, linearity, interference, and storage stability. Method comparison used an immunoturbidimetric comparator assay (Olympus DxC 700 AU). Diagnostic performance was evaluated via receiver operating characteristic (ROC) analysis, likelihood ratios, and odds ratios.

RESULTS: Average intra- and inter-assay coefficients of variation were 3.92% and 8.47%. Linearity over the clinical range (0.86%-6.27% HbA1c%) was strong (R2 = 0.973) with minor constant bias. HbA1c remained stable for 5 days at 4 °C and up to 3 months at both -20 and -80 °C, but not reliably at 6 months. Lipemia and icterus exhibited variable interfering effects. Method comparison indicated non-interchangeability with the comparator assay. Diabetic dogs had significantly higher HbA1c than NDM (mean 5.10% vs 1.58%). ROC analysis yielded an AUC of 0.969, with a cutoff > 3.22% achieving 90.20% sensitivity and 100% specificity. The positive likelihood ratio was not estimable, and the odds ratio was 22.90 (95% CI, 3.60-146.90). Healthy young dogs' reference interval was 1.25% (90% CI, 1.19-1.59) to 2.75% (90% CI, 2.56-2.90).

CONCLUSIONS AND CLINICAL IMPORTANCE: The Vet Chroma™ provides reliable HbA1c measurements in canine whole blood and effectively discriminates diabetic from non-diabetic dogs. Results should be interpreted alongside clinical and clinicopathologic findings.