MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.athoracsur.2011.01.042

Медицинская библиотека · The Annals of thoracic surgery · 2011

Intermediate and early rupture of expanded polytetrafluoroethylene neochordae after mitral valve repair.

Yamashita Michael H, Skarsgard Peter L

СтатьяПеревод в очереди

Аннотация

Neochordae construction using an expanded polytetrafluoroethylene suture is an accepted surgical technique to correct mitral regurgitation due to mitral valve prolapse. We report two cases of intermediate and early rupture of CV-5 expanded polytetrafluoroethylene neochordae after mitral valve repair. As a result of these cases, we now use CV-3 expanded polytetrafluoroethylene for neochordae construction with no failures.