MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hroo.2025.11.002

Медицинская библиотека · Heart rhythm O2 · 2025

Inappropriate shocks from subcutaneous and transvenous implantable cardioverter-defibrillators: Reports from the FDA MAUDE database.

Hauser Robert G, Desouki Mariam, Stanberry Larissa I, Abdelsayed Kerollos, Kapphahn-Bergs Melanie, Witt Dawn, Sengupta Jay N

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The subcutaneous implantable cardioverter-defibrillator (S-ICD) has been shown to be safe and effective for preventing sudden cardiac death. Clinical trial and registry data suggest that the incidence of S-ICD inappropriate shocks (IAS) is similar to or slightly higher than IAS from transvenous implantable cardioverter-defibrillators. However, little data exist for S-ICD IAS in everyday practice.

OBJECTIVE

This study aimed to assess IAS events reported by Boston Scientific (BSC) to the Food and Drug Administration (FDA) for its Emblem S-ICDs and transvenous single-chamber implantable cardioverter-defibrillators (TV-VR ICDs).

METHODS

We searched for reports of Emblem S-ICD and TV-VR ICD adverse events (AEs) in the FDA's Manufacturer and User Facility Device Experience (MAUDE) database that were received by the FDA from BSC during 2020-2024.

RESULTS

In 2024, BSC estimated 153,000 Emblem S-ICDs and 353,000 of its TV-VR ICDs were distributed worldwide. During 2020-2024, BSC submitted 22,957 MAUDE AE reports for Emblem S-ICDs and 6791 MAUDE AE reports for TV-VR ICDs. Of these, 5363 (23.4%) involved IAS for S-ICDs and 813 IAS (11.7%) for TV-VR ICDs. The most common cause of S-ICD IAS was oversensing of physiological and nonphysiological signals (91.1%; 4885 of 5363), whereas rapid atrial fibrillation or supraventricular tachycardia caused 69.5% of TV-VR ICD IAS. Emblem S-ICD IAS induced ventricular fibrillation or ventricular tachycardia in 69 patients and failed to defibrillate/convert 4 of them.

CONCLUSION

These real-word data suggest that S-ICD IAS may be more common than previously reported. The incidence, root causes, and consequences of these IAS in the general population of S-ICD recipients should be investigated.

Для цитирования: Hauser Robert G, Desouki Mariam, Stanberry Larissa I, Abdelsayed Kerollos, Kapphahn-Bergs Melanie, Witt Dawn, Sengupta Jay N. Inappropriate shocks from subcutaneous and transvenous implantable cardioverter-defibrillators: Reports from the FDA MAUDE database.. Heart rhythm O2. 2025. DOI: 10.1016/j.hroo.2025.11.002.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.