BACKGROUND: Fractional flow reserve (FFR)-guided revascularization strategy is popular in coronary intervention. However, the feasibility of assessing stenotic severity in intracranial large arteries using pressure gradient measurements still remains unclear.
METHODS: Between March 2013 and May 2014, 12 consecutive patients with intracranial large artery stenosis (including intracranial internal carotid artery, middle cerebral M1 segment, intracranial vertebral artery, and basilar artery) were enrolled in this study. The trans-stenotic pressure gradient was measured before and/or after percutaneous transluminal angioplasty and stenting (PTAS), and was then compared with percent diameter stenosis. A Pd /Pa cut-off of ≤0.70 was used to guide stenting of hemodynamically significant stenoses. The device-related and procedure-related serious adverse events and recurrent cerebral ischemic events were recorded.
RESULTS: The target vessel could be reached in all cases. No technical complications occurred due to the specific study protocol. Excellent pressure signals were obtained in all patients. For seven patients who performed PTAS, the mean pre-procedural pressure gradient decreased from 59.0 ± 17.2 to 13.3 ± 13.6 mm Hg after the procedure (P < 0.01). Only one patient who refused stenting experienced a TIA event in the ipsilateral MCA territory. No recurrent ischemic event was observed in other patients.
CONCLUSION: Mean trans-stenotic pressure gradients can be safely and easily measured with a 0.014-inch fluid-filled guide wire in intracranial large arteries. © 2016 Wiley Periodicals, Inc.
METHODS: Between March 2013 and May 2014, 12 consecutive patients with intracranial large artery stenosis (including intracranial internal carotid artery, middle cerebral M1 segment, intracranial vertebral artery, and basilar artery) were enrolled in this study. The trans-stenotic pressure gradient was measured before and/or after percutaneous transluminal angioplasty and stenting (PTAS), and was then compared with percent diameter stenosis. A Pd /Pa cut-off of ≤0.70 was used to guide stenting of hemodynamically significant stenoses. The device-related and procedure-related serious adverse events and recurrent cerebral ischemic events were recorded.
RESULTS: The target vessel could be reached in all cases. No technical complications occurred due to the specific study protocol. Excellent pressure signals were obtained in all patients. For seven patients who performed PTAS, the mean pre-procedural pressure gradient decreased from 59.0 ± 17.2 to 13.3 ± 13.6 mm Hg after the procedure (P < 0.01). Only one patient who refused stenting experienced a TIA event in the ipsilateral MCA territory. No recurrent ischemic event was observed in other patients.
CONCLUSION: Mean trans-stenotic pressure gradients can be safely and easily measured with a 0.014-inch fluid-filled guide wire in intracranial large arteries. © 2016 Wiley Periodicals, Inc.
Предметные рубрики MeSH
AdultAgedAngioplasty, BalloonArterial Occlusive DiseasesArterial PressureBasilar ArteryBlood Pressure DeterminationCarotid Artery, InternalCerebral AngiographyConstriction, PathologicEquipment DesignFeasibility StudiesFemaleHumansIntracranial Arterial DiseasesMaleMiddle AgedMiddle Cerebral ArteryPredictive Value of TestsReproducibility of ResultsSeverity of Illness IndexStentsTransducers, PressureTreatment OutcomeVertebral Artery