DOI 10.1016/j.jcrs.2014.02.032

Медицинская библиотека · Journal of cataract and refractive surgery · 2014

Deformation response of paired donor corneas to an air puff: intact whole globe versus mounted corneoscleral rim.

Metzler Kimberly M, Mahmoud Ashraf M, Liu Jun, Roberts Cynthia J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate the influence of ocular shell biomechanical characteristics on corneal deformation response to an air puff.

SETTING

The Ohio State University, Columbus, Ohio, USA.

DESIGN

Experimental study.

METHODS

Twenty-four eyes of 12 human donors were obtained in matched pairs. One eye was secured in a purpose-designed whole globe mount (whole-globe group). The cornea from the fellow eye was placed in a Barron artificial anterior chamber (artificial-chamber group). The corneas were mounted sequentially and connected to a pressure-control system. Deformation data were acquired using the Corvis ST, a dynamic Scheimpflug analyzer. Internal pressure was set to 10, 20, 30, 40, and 50 mm Hg; at least 4 examinations were performed at each pressure.

RESULTS

Statistically significantly higher maximum deformation amplitude was observed in the whole-globe group than in the artificial-chamber group at all pressures. The mean amplitude differences were 1.006 mm ± 0.238 [SD], 0.614 ± 0.137 mm, 0.384 ± 0.099 mm, 0.229 ± 0.087 mm, and 0.133 ± 0.068 mm at 10, 20, 30, 40, and 50 mm Hg, respectively (P<.0001, P<.0001, P<.0001, P<.0001, and P<.0002, respectively). Nonlinear regression of the deformation amplitude differences between pairs showed a significant decrease with increasing pressure (P<.0001, R(2) = 0.8385).

CONCLUSIONS

The deformation response to an air puff was affected by the type of mount used, with a stiffer shell producing a stiffer corneal response and decreasing differences at higher internal pressures. In vivo air-puff examinations may be affected by scleral stiffness in addition to the cornea.

FINANCIAL DISCLOSURES: Dr. Roberts is a consultant to Oculus Optikgeräte GmbH and Ziemer Ophthalmic Systems AG and has received research funding from Carl Zeiss Meditec AG and travel funds from Sooft Italia. No other author has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Metzler Kimberly M, Mahmoud Ashraf M, Liu Jun, Roberts Cynthia J. Deformation response of paired donor corneas to an air puff: intact whole globe versus mounted corneoscleral rim.. Journal of cataract and refractive surgery. 2014. DOI: 10.1016/j.jcrs.2014.02.032.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.