MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.rpth.2026.106636

Медицинская библиотека · Research and practice in thrombosis and haemostasis · 2026

Resolving the dilution paradox to improve the interpretation of extracellular vesicle biomarker studies.

Rops Joyce, Buntsma Naomi C, Gąsecka Aleksandra, Nieuwland Rienk, Rosiek Aleksandra, Roos Yvo B W E M, Nossent A Yaël, Kruyt Nyika D, van der Pol Edwin, Eyileten Ceren

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Concentrations of extracellular vesicles (EVs) and other particles are measured in plasma for biomarker exploration. A commonly used method, flow cytometry, requires plasma dilution to ensure single-particle detection. Since plasma EVs are outnumbered by variable concentrations of lipoproteins, dilution differs between the plasma samples. Dilution can result in misidentification of fluorescent background signals as labeled EVs. This phenomenon, called the dilution paradox, leads to overestimation of plasma EV concentrations, and likely impacts conclusions from earlier performed biomarker studies.

OBJECTIVES

This study reevaluated earlier conclusions from our clinical biomarker studies Antiplatelet Therapy Effect on Extracellular Vesicles (AFFECT EV) and Circulating Nanotraces to Identify the Cause of Stroke (CINTICS), by taking the dilution paradox into account.

METHODS

We developed a model that quantifies the fluorescent background and estimates whether a flow cytometry measurement is dominated by fluorescent background, that is, if the measurement is unreliable. This model was applied to the original datasets of the AFFECT EV and CINTICS studies to identify and exclude unreliable measurements. We investigated whether exclusion of unreliable data affects the original conclusions.

RESULTS

Our model estimated that 47% (1156/2457) of the evaluated measurements are unreliable, and conclusions from both biomarker studies required adjustment.

CONCLUSION

Our model improves reliability and reproducibility of EV concentration measurements using flow cytometry. We recommend to reanalyze earlier EV flow cytometry studies using our model and to use a fixed dilution factor in future EV flow cytometry studies to enable reliable EV concentration measurements.

Для цитирования: Rops Joyce, Buntsma Naomi C, Gąsecka Aleksandra, Nieuwland Rienk, Rosiek Aleksandra, Roos Yvo B W E M, Nossent A Yaël, Kruyt Nyika D, van der Pol Edwin, Eyileten Ceren. Resolving the dilution paradox to improve the interpretation of extracellular vesicle biomarker studies.. Research and practice in thrombosis and haemostasis. 2026. DOI: 10.1016/j.rpth.2026.106636.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.