Safety and Efficacy of Hypersonic Vitrectomy for Retinal Reattachment Surgery in Proliferative Vitreoretinopathies.

Deuchler Svenja, Scholtz Julia, Knoch Timo, Seitz Berthold, Koch Frank H

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

To investigate the safety and efficacy of the Vitesse hypersonic vitrectomy device for retinal reattachment surgery in proliferative vitreoretinopathy (PVR) or proliferative diabetic vitreoretinopathy (PDVR) cases. The Vitesse device utilizes hypersonic technology to liquefy instead of cutting the vitreous, providing an alternative to the traditional pneumatic guillotine cutter.

MATERIAL AND

METHODS

A prospective, one-armed, non-comparative, open-label study was performed. Sixteen patients with a diagnosis of PVR or PDVR that required retinal reattachment surgery were included. Severity of disease was classified using the Retina Society 1983 classification and Kroll Classification from 2007. Patient data was collected preoperatively, 2 days postoperatively, 1 month postoperatively, and 3 months postoperatively. Efficacy of hypersonic vitrectomy was evaluated, both subjectively using a questionnaire and objectively by means of Supplementary Video documentation, device settings, and data collection of the patients' medical history.

RESULTS

In all 16 cases, retinal reattachment surgery could be performed with Vitesse without conversion to a guillotine cutter. The vitreous could be separated from the detached retina completely with no iatrogenic tissue damage. Vitreous traction was documented in one case. In 6 cases, shaving of the vitreous base was performed after early fluid-air exchange with hypersonic vitrectomy without any complications. No adverse events suspected to be related to hypersonic vitrectomy were documented in the follow-up visits.

CONCLUSION

The Vitesse has potential advantages including reduced vacuum volumes with limited amount of turbulence and avoidance vitreoretinal traction. Liquefaction directly in front of the probe entrance allows for continuous unrestricted fluid flow. These factors of hypersonic technology allow to carry out retinal reattachment surgery in PVR or PDVR cases successfully.

Для цитирования: Deuchler Svenja, Scholtz Julia, Knoch Timo, Seitz Berthold, Koch Frank H. Safety and Efficacy of Hypersonic Vitrectomy for Retinal Reattachment Surgery in Proliferative Vitreoretinopathies.. Clinical ophthalmology (Auckland, N.Z.). 2022. DOI: 10.2147/opth.s386014.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.