BACKGROUND

We investigated the incidence of embolic protection device retrieval difficulties at carotid artery stenting (CAS) with a closed-cell stent and demonstrated the usefulness of a manual carotid compression assist technique.

METHODS

Between July 2010 and October 2013, we performed 156 CAS procedures using self-expandable closed-cell stents. All procedures were performed with the aid of a filter design embolic protection device. We used FilterWire EZ in 118 procedures and SpiderFX in 38 procedures. The embolic protection device was usually retrieved by the accessory retrieval sheath after CAS. We applied a manual carotid compression technique when it was difficult to navigate the retrieval sheath through the deployed stent. We compared clinical outcomes in patients where simple retrieval was possible with patients where the manual carotid compression assisted technique was used for retrieval.

RESULTS

Among the 156 CAS procedures, we encountered 12 (7.7%) where embolic protection device retrieval was hampered at the proximal stent terminus. Our manual carotid compression technique overcame this difficulty without eliciting neurologic events, artery dissection, or stent deformity.

CONCLUSIONS

In patients undergoing closed-cell stent placement, embolic protection device retrieval difficulties may be encountered at the proximal stent terminus. Manual carotid compression assisted retrieval is an easy, readily available solution to overcome these difficulties.

Для цитирования: Nii Kouhei, Nakai Kanji, Tsutsumi Masanori, Aikawa Hiroshi, Iko Minoru, Sakamoto Kimiya, Mitsutake Takafumi, Eto Ayumu, Hanada Hayatsura, Kazekawa Kiyoshi. A manual carotid compression technique to overcome difficult filter protection device retrieval during carotid artery stenting.. Journal of stroke and cerebrovascular diseases : the official journal of National Stroke Association. 2014. DOI: 10.1016/j.jstrokecerebrovasdis.2014.08.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.