Safety and Efficacy of Magnetically Controlled Growing Rods for Early Onset Scoliosis in the Era of Disease-Modifying Therapies: Mid-term Results in a Homogeneous Cohort of Spinal Muscular Atrophy Type 1 Patients.

Labianca Luca, Calderaro Cosma, Manca Maria Beatrice, Manni Filippo, Colantoni Edoardo, Benelli Cristiano, Maffulli Nicola

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The advent of disease-modifying therapies (DMTs) has fundamentally altered the clinical course of spinal muscular atrophy type 1 (SMA-I), resulting in prolonged survival and a higher incidence of early onset scoliosis (EOS). Although growth-friendly surgery is a keystone of management, specific evidence regarding magnetically controlled growing rods (MCGR) in the SMA-I population remains sparse. We evaluate the safety and efficacy of MCGR in a specific, homogeneous cohort of SMA-I patients.

METHODS

A retrospective investigation was performed on 8 consecutive children with genetically confirmed SMA-I who underwent MCGR implantation for progressive EOS at our institution. To overcome profound osteopenia, we utilized a proximal rib-based "ring hook" foundation and distal pelvic/sacroiliac fixation. Outcomes included coronal Cobb angle correction, pelvic alignment, and annualized T1-S1 spinal growth. Distractions were performed at 10-12-week intervals and validated via ultrasonography. Complications and unplanned returns to the operating room (UPRORs) were recorded.

RESULTS

Eight patients (mean age 6.9 ± 0.92 years) were followed for a mean of 29.25 ± 4.0 months. The mean Cobb angle improved from 71.9° ± 20.8° to 38.9° ± 15.7° postoperatively (45.9% initial correction; P < .001) and was 44.5° ± 14.1° at the final follow-up. Pelvic obliquity significantly improved from a mean of 25.6° ± 6.3° to 6.6° ± 3.0° (P < .001) and was at the last follow-up (8.7° ± 2.0°; P = .08). Annualized T1-S1 growth was 11.6 ± 3.2 mm/y, with a mean "distraction efficiency" of 86.7% compared to the programmed 3 mm target. One mechanical complication (12.5%) occurred (distal iliac screw failure), resulting in a UPROR rate of 6.25 per 100 patient-years. No perioperative respiratory or infectious complications were observed.

CONCLUSIONS

In this medically fragile SMA-I cohort, MCGR provided effective deformity control and supported longitudinal spinal growth. A multidisciplinary perioperative pathway and rib-based proximal anchoring are essential for optimizing results. While mid-term outcomes are encouraging, further research is needed to evaluate the long-term impact on pulmonary function and construct durability in the post-disease-modifying therapy era.

KEY CONCEPTS: (1)Meaningful deformity correction (45.9% mean Cobb reduction) achieved with preservation of linear spinal growth (11.6 ± 3.2 mm/y).(2)Low perioperative risk, with no infections, no transfusions, and no perioperative respiratory deterioration.(3)A 12.5% mechanical complication rate (one unplanned revision for distal iliac screw failure) over a mean follow-up of 29.25 months.

LEVEL OF EVIDENCE: IV: Case series.

Для цитирования: Labianca Luca, Calderaro Cosma, Manca Maria Beatrice, Manni Filippo, Colantoni Edoardo, Benelli Cristiano, Maffulli Nicola. Safety and Efficacy of Magnetically Controlled Growing Rods for Early Onset Scoliosis in the Era of Disease-Modifying Therapies: Mid-term Results in a Homogeneous Cohort of Spinal Muscular Atrophy Type 1 Patients.. Journal of the Pediatric Orthopaedic Society of North America. 2026. DOI: 10.1016/j.jposna.2026.100368.