Real-time magnetic resonance-guided radiofrequency ablation and lesion evaluation in an magnetic resonance-compatible isolated beating pig heart platform.

Hopman Luuk H G A, Schrauben Eric M, Nelissen Jules L, Reitzema Pieter J, Boonstra Machteld J, Smeets Bart L M, Boekhoven Renate W, Sunnarborg Daniel, Smink Jouke, Niessen Hans W M, Allaart Cornelis P, Nederveen Aart J, Götte Marco J W

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Interventional cardiovascular magnetic resonance imaging (MRI) offers real-time, radiation-free guidance for complex procedures such as myocardial ablation, marking a promising advance in electrophysiology. However, further development is limited by challenges in magnetic resonance (MR)-compatible instrument testing, MRI sequence validation, and accurate correlation with histopathology, hindered by the limitations of in vivo tissue evaluation.

OBJECTIVE

This study investigated the feasibility of real-time MR-guided radiofrequency (RF) ablation in an MR-compatible isolated beating pig heart platform and characterized ablation lesions using MRI and histopathology.

METHODS

A heart from a pig slaughtered for human consumption was prepared under regulatory guidelines and connected to a custom-built, MR-compatible perfusion platform supporting left ventricular function in both Langendorff and working modes. Autologous heparinized blood circulated at physiological pressures and temperatures. MR-guided catheter navigation and RF ablation were performed on a Philips 3T scanner using active catheter tracking. Native T1 and T2 mapping were acquired before and after ablation. Lesions were confirmed by histologic analysis.

RESULTS

RF ablation (50 W, 60 seconds) was successfully performed at 5 left ventricular sites. MRI showed focal reductions in T1 (936 ± 80 ms) surrounded by elevated T1 (1357 ± 18 ms) and T2 values (86 ± 10 ms) compared with nonablated myocardium (T1 1192 ± 26 ms; T2 66 ± 6 ms), consistent with necrosis and edema. Histology confirmed a necrotic core with a surrounding rim showing contraction band necrosis and erythrocyte extravasation.

CONCLUSION

This study demonstrates the feasibility of real-time MR-guided ablation in a beating pig heart platform. The setup allows high-resolution lesion assessment and histologic correlation, supporting future developments in MR-guided therapies.

Для цитирования: Hopman Luuk H G A, Schrauben Eric M, Nelissen Jules L, Reitzema Pieter J, Boonstra Machteld J, Smeets Bart L M, Boekhoven Renate W, Sunnarborg Daniel, Smink Jouke, Niessen Hans W M, Allaart Cornelis P, Nederveen Aart J, Götte Marco J W. Real-time magnetic resonance-guided radiofrequency ablation and lesion evaluation in an magnetic resonance-compatible isolated beating pig heart platform.. Heart rhythm O2. 2025. DOI: 10.1016/j.hroo.2025.09.014.