MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2019.07.058

Медицинская библиотека · JACC. Cardiovascular interventions · 2019

Safety and Performance Outcomes of a Self-Expanding Transcatheter Aortic Heart Valve: The BIOVALVE Trials.

Schäfer Ulrich, Kempfert Jörg, Verheye Stefan, Maisano Francesco, Thiele Holger, Landt Martin, Haude Michael, Rudolph Tanja K, Ince Hüseyin, Kische Stephan, Treede Hendrik, Tonino Pim, Conradi Lenard

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The aim of this study was to assess the safety and performance of a newly introduced transfemoral self-expanding supra-annular transcatheter heart valve.

BACKGROUND

Transcatheter aortic valve replacement has become an established procedure, and newer transcatheter heart valves aim to supplement the current armamentarium.

METHODS

BIOVALVE-I is a prospective, single-center feasibility study enrolling 13 high-risk patients with severe aortic stenosis, and BIOVALVE-II is a multicenter pilot study enrolling a further 55 patients under the same protocol.

RESULTS

Patients' mean age was 82.4 ± 5.3 years, 57.4% had concomitant coronary artery disease, 50.0% had renal insufficiency stage ≥3, 32.4% had diabetes mellitus, 29.4% had current anemia, 19.1% had chronic obstructive pulmonary disease, 19.1% were frail, and 17.6% had prior cerebrovascular events. The primary endpoint, early safety according to the Valve Academic Research Consortium-2 guidelines, was observed in 13 patients (19.1%). One patient (1.5%) died through 30 days of a noncardiac cause, and 4 patients (7.8%) died through 180 days. Disabling stroke occurred at 30 days in 1 patient (1.5%) and at 180 days in 2 patients (3.2%). New permanent pacemakers up to 30 days were implanted in 9 patients (13.4%). Two patients (3.8%) had moderate or severe paravalvular leakage at 30 days and 4 (9.1%) at 6 months, mean effective orifice area was 2.0 ± 0.4 cm2 and 1.9 ± 0.5 cm2, and mean gradient was 6.4 mm Hg at 30 days and 6 months.

CONCLUSIONS

Clinical outcomes with the Biovalve self-expanding transcatheter heart valve were similar to those seen in other first-in-human studies with first-generation devices for severe aortic stenosis.

Для цитирования: Schäfer Ulrich, Kempfert Jörg, Verheye Stefan, Maisano Francesco, Thiele Holger, Landt Martin, Haude Michael, Rudolph Tanja K, Ince Hüseyin, Kische Stephan, Treede Hendrik, Tonino Pim, Conradi Lenard. Safety and Performance Outcomes of a Self-Expanding Transcatheter Aortic Heart Valve: The BIOVALVE Trials.. JACC. Cardiovascular interventions. 2019. DOI: 10.1016/j.jcin.2019.07.058.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.