World neurosurgery · 2021

Comaneci Device for Temporary Coiling Assistance for Treatment of Wide-Necked Aneurysms: Initial Case Series and Systematic Literature Review.

СтатьяСистематический обзорНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Lim Jaims, Vakharia Kunal, Waqas Muhammad, Affana Clementine, Siddiqui Adnan H, Davies Jason M, Levy Elad I

Аннотация

BACKGROUND: Coiling of wide-necked aneurysms requires high-density packing of coils within the aneurysm, which necessitates adequate microcatheter access and navigability. The Comaneci device, introduced in the United States in 2019, is a retrievable stent that can be used as an adjunct to coiling of a wide-necked aneurysm without limiting flow. We present a case series and systematic review of use of this device.

METHODS: All cases involving use of the device at our institute between May 1, 2019, and April 30, 2020, were reviewed. A comprehensive systematic review of the literature was conducted using PubMed and EMBASE and a review of eligible article bibliographies.

RESULTS: Five patients underwent Comaneci-assisted wide-necked aneurysm coiling during the study period; 4 were treated via a radial artery approach, and 1 was treated via a femoral artery approach. Two patients presented with subarachnoid hemorrhage: 1 had a ruptured posterior inferior cerebellar artery aneurysm (Hunt-Hess 5, Fisher 4), and 1 had a ruptured middle cerebral artery aneurysm (Hunt-Hess 2, Fisher 1). Mean aneurysmal neck size was 4.4 ± 0.8mm; mean aspect ratio was 1.2 ± 0.3. Raymond-Roy 1 occlusion was achieved in all aneurysms except the posterior inferior cerebellar artery aneurysm. Systematic literature review identified 4 articles that found use of the Comaneci device to treat wide-necked aneurysms to be effective.

CONCLUSIONS: This device can be used with transfemoral and transradial approaches, allowing for continued flow through the parent vessel during the coiling procedure while providing a scaffold for dense coiling of the aneurysm and its neck.