MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.ijg.0000138205.99424.24

Медицинская библиотека · Journal of glaucoma · 2004

Evaluation of scanning resolution on retinal nerve fiber layer measurement using optical coherence tomography in normal and glaucomatous eyes.

Leung C K S, Yung W H, Ng A C K, Woo J, Tsang M K, Tse K K

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the effect of varying the scanning resolution of optical coherence tomography (OCT) retinal nerve fiber layer (RNFL) measurement on diagnostic sensitivity and functional correlation in glaucoma.

PATIENTS AND

METHODS

314 eyes from 182 subjects including 107 normal eyes, 83 glaucoma suspect eyes, and 124 glaucoma eyes were included in this cross-sectional study. Standard automated perimetry and OCT measurement of RNFL thickness were performed. Each individual underwent two scanning protocols: (1) fast RNFL thickness (3.4) scan (with resolution of 256 scan points) and (2) RNFL thickness (3.4) scan (with resolution of 512 scan points). RNFL thickness was compared among the groups. Diagnostic sensitivity was evaluated with Receiver Operating Characteristic (ROC) Curve. Relationship between RNFL thickness and visual field mean deviation was examined using linear regression analysis.

RESULTS

Measured RNFL thickness using fast RNFL thickness (3.4) scan was significantly higher compared with RNFL thickness (3.4) scan in average, superior, nasal and inferior RNFL in all diagnostic groups. Comparing normal and glaucoma groups, RNFL thickness (3.4) scan produced the largest area under the ROC curve (0.912) based on average RNFL thickness. A stronger correlation between average RNFL and visual field mean deviation was found in RNFL thickness (3.4) scan (R = 0.75, R = 0.56).

CONCLUSIONS

Higher resolution RNFL scan provides better diagnostic sensitivity in glaucoma detection and a stronger correlation with visual function.

Для цитирования: Leung C K S, Yung W H, Ng A C K, Woo J, Tsang M K, Tse K K. Evaluation of scanning resolution on retinal nerve fiber layer measurement using optical coherence tomography in normal and glaucomatous eyes.. Journal of glaucoma. 2004. DOI: 10.1097/01.ijg.0000138205.99424.24.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.