MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-2008-1050444

Медицинская библиотека · Klinische Monatsblatter fur Augenheilkunde · 1987

[Scanning electron microscopy findings in a rigid one-piece polymethylmethacrylate anterior chamber lens with a 30 year duration in the eye].

Rochels R, Stofft E

Клинический случайАннотация на английскомСтатьяПеревод в очереди

Аннотация

A rigid one-piece PMMA anterior chamber lens (Scharf type) had to be removed 30 years after implantation because of diffuse bullous keratopathy and painful secondary glaucoma. Scanning electron microscopy showed the IOL to have a crater-like surface, while the haptics, having been in contact with intraocular tissues, were superficially and deeply eroded. Comparison with two identical anterior chamber lenses which had not been implanted revealed that the changes seen in the removed IOL must have been the result of degradation of the PMMA material.

Для цитирования: Rochels R, Stofft E. [Scanning electron microscopy findings in a rigid one-piece polymethylmethacrylate anterior chamber lens with a 30 year duration in the eye].. Klinische Monatsblatter fur Augenheilkunde. 1987. DOI: 10.1055/s-2008-1050444.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.