DOI 10.1097/j.jcrs.0000000000001904
Agreement between a new SS-OCT biometer and a SD-OCT/Placido topographer for anterior segment measurements.
Аннотация
PURPOSE: To evaluate the agreement of anterior segment biometric measurements between a new swept-source optical coherence tomography (SS-OCT) biometer (ZW-30) and a spectral-domain optical coherence tomography combined with Placido-disk topography (SD-OCT/Placido topographer, MS-39).
SETTING
Eye and ENT Hospital of Fudan University, Shanghai, China.
DESIGN
Prospective cross-sectional study.
METHODS
This prospective, cross-sectional study recruited 70 healthy eyes from 70 participants. Each eye underwent 3 consecutive scans with both ZW-30 and MS-39. Parameters analyzed included central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), J 0 and J 45 astigmatism vector components, and corneal diameter (CD). Interdevice differences were evaluated using paired t tests, and Bland-Altman analysis with 95% limits of agreement (LoA) assessed the agreement.
RESULTS
Significant differences were observed in all parameters measured, except for J 0 . Bland-Altman analysis indicated high agreement for CCT (mean difference: 5.38 ± 3.67 μm; 95% LoA: -2.73 to 13.48 μm), ACD (-0.11 ± 0.03 mm; -0.18 to -0.03 mm), LT (-0.06 ± 0.05 mm; -0.17 to 0.05 mm), Km (0.07 ± 0.16 diopter [D]; -0.33 to 0.47 D), J 0 (0.02 ± 0.12 D; -0.31 to 0.27 D), and J 45 (-0.08 ± 0.09 D; -0.32 to 0.16 D). A larger discrepancy was observed for CD (0.48 ± 0.13 mm; 0.17 to 0.79 mm).
CONCLUSIONS
The ZW-30 SS-OCT biometer and the MS-39 SD-OCT/Placido topographer demonstrated high agreement for most anterior segment parameters, indicating their interchangeability in clinical practice. However, caution is advised when using CD measurements interchangeably because of the observed systematic differences between devices.