Safety and efficacy of low-energy small incision lenticule extraction for the correction of myopia and myopic astigmatism: A retrospective analysis.
Аннотация
PURPOSE: To determine the safety and efficacy of low-energy settings in small incision lenticule extraction (SMILE) for correcting myopia and myopic astigmatism.
METHODS
We included patients aged ≥18 years with the myopia of -0.5 to -10 D and myopic astigmatism of -0.5 to -5 Dcyl in this retrospective case series performed at a private eye hospital in South India. All patients had preoperative best-corrected visual acuity of LogMar 0.0 ± 0, with stable refraction for 1 year and normal corneal topography. Ocular surface disease and other pathology cases were excluded. The repetition rate of the laser was 500 kHz, and the pulse energy was 110 nJ. The lenticule diameter was set at 6.5 mm, cap diameter was 7.20 mm, and intended cap thickness was 110-130 μ. The spot distance was 4.5 μm. All patients were evaluated immediately postoperation and on postoperative days 1, 8, and 30.
RESULTS
Overall, 541 eyes were included. The mean patient age was 25.03 ± 4.1 years. The mean spherical error was -3.76 ± 1.84 Ds. The mean cylinder was -1.24 ± 0.91. The mean spherical equivalent of refraction was -4.22 ±1.94 D. The logMAR on postoperative day 1 was 0.0 ± 0. The mean spherical equivalent at 1 month was 0.28 ± 1.06 D. There was no loss of Snellen's lines after the procedure. The mean spherical equivalent of refraction to the target was 95% within ± 0.50 D. The postoperative astigmatism was within 0.5 Dycl. No intraoperative complications of SMILE including retained lenticule fragments, tears of incision, or improper dissection occurred.
CONCLUSION
Low-energy settings in SMILE are safe and effective in correcting myopia and myopic astigmatism including high cylinders (>3 Dcyl).
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.