Evaluation of Long-term Rotational Stability and Functional Performance of Two Toric Multifocal IOL Models Beyond 5 Years.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Köppe Maximilian K, Auffarth Gerd U, Son Hyeck-Soo, Yildirim Timur M, Khoramnia Ramin
Аннотация
PURPOSE: To assess visual outcomes and stability of lens alignment more than 3 years after implanting either the AT Lisa Tri Toric (Carl Zeiss Meditec) or the Acrysof IQ Restor Toric (Alcon Laboratories, Inc) intraocular lens (IOL).
METHODS: Patients who had received either the AT Lisa Tri Toric or the Acrysof IQ Restor Toric IOL were included in the study. The average follow-up period was 5.4 years for the AT Lisa Tri Toric group and 7.9 years for the Acrysof IQ Restor group. The examinations covered included: subjective refraction, monocular and binocular corrected and uncorrected visual acuity at far (4 m), intermediate (75 cm), and near (40 cm) distances, binocular best-corrected defocus curve analysis (4 m), contrast sensitivity evaluation, assessment of IOL axial position, and presence of positive dysphotopsia.
RESULTS: A total of 42 eyes were evaluated. Both groups exhibited good long-term lens axial stability. The AT Lisa Tri Toric group had a mean misalignment of 4.8 ± 3.2° from the preoperatively determined axis, whereas the Acrysof IQ Restor Toric group had an average misalignment of 5.4 ± 4.5° (no statistically significant difference, P = .578). Mean achieved spherical equivalent was -0.19 ± 0.36 diopters (D) for the AT Lisa Tri Toric group and 0.15 ± 0.39 D for the Acrysof IQ Restor Toric group. In the AT Lisa Tri Toric group, 18 eyes (72%) achieved an uncorrected distance visual acuity of 0.10 logarithm of the minimum angle of resolution (20/25 Snellen) or better, compared to 15 eyes (88%) in the Acrysof IQ Restor Toric group. Overall, 21 eyes (84%) in the AT Lisa Tri Toric group and 14 eyes (83%) in the Acrysof IQ Restor Toric group achieved a postoperative spherical equivalent refraction between -0.50 and +0.50 D.
CONCLUSIONS: Both the AT Lisa Tri Toric and the Acrysof IQ Restor Toric IOLs showed good long-term stability and functional outcomes. They proved to be effective options for restoring vision at far, intermediate, and near distances, providing a viable solution for presbyopia correction even several years after implantation. [J Refract Surg. 2024;40(11):e845-e853.].
METHODS: Patients who had received either the AT Lisa Tri Toric or the Acrysof IQ Restor Toric IOL were included in the study. The average follow-up period was 5.4 years for the AT Lisa Tri Toric group and 7.9 years for the Acrysof IQ Restor group. The examinations covered included: subjective refraction, monocular and binocular corrected and uncorrected visual acuity at far (4 m), intermediate (75 cm), and near (40 cm) distances, binocular best-corrected defocus curve analysis (4 m), contrast sensitivity evaluation, assessment of IOL axial position, and presence of positive dysphotopsia.
RESULTS: A total of 42 eyes were evaluated. Both groups exhibited good long-term lens axial stability. The AT Lisa Tri Toric group had a mean misalignment of 4.8 ± 3.2° from the preoperatively determined axis, whereas the Acrysof IQ Restor Toric group had an average misalignment of 5.4 ± 4.5° (no statistically significant difference, P = .578). Mean achieved spherical equivalent was -0.19 ± 0.36 diopters (D) for the AT Lisa Tri Toric group and 0.15 ± 0.39 D for the Acrysof IQ Restor Toric group. In the AT Lisa Tri Toric group, 18 eyes (72%) achieved an uncorrected distance visual acuity of 0.10 logarithm of the minimum angle of resolution (20/25 Snellen) or better, compared to 15 eyes (88%) in the Acrysof IQ Restor Toric group. Overall, 21 eyes (84%) in the AT Lisa Tri Toric group and 14 eyes (83%) in the Acrysof IQ Restor Toric group achieved a postoperative spherical equivalent refraction between -0.50 and +0.50 D.
CONCLUSIONS: Both the AT Lisa Tri Toric and the Acrysof IQ Restor Toric IOLs showed good long-term stability and functional outcomes. They proved to be effective options for restoring vision at far, intermediate, and near distances, providing a viable solution for presbyopia correction even several years after implantation. [J Refract Surg. 2024;40(11):e845-e853.].