MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/icvts/ivab101

Медицинская библиотека · Interactive cardiovascular and thoracic surgery · 2021

Simultaneous transcatheter treatment of ascending aortic aneurysm with aortic and mitral regurgitation: an in vitro study.

Zhao Yuxi, Li Tao, Wu Mingwei, Zeng Zhaoxiang, Gao Minxin, Bao Xianhao, Xu Ziyi, Zhou Jian, Feng Jiaxuan, Feng Rui, Jing Zaiping

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

We sought to explore the efficacy of the endovascular repair of an ascending aortic aneurysm with aortic and mitral regurgitation by 2 novel valved stents.

METHODS

We established models of ascending aortic aneurysms combined with aortic and mitral regurgitation in 10 pig hearts, then implanted self-expanding aortic fenestrated and mitral valved stents via the transapical approach. We applied a fluoroscopy-guided in vitro setting to test the approach, then analysed continuous circulating flushing at 37°C. Finally, we determined operating times, echocardiography and changes of coronary flow as well as fenestration alignment with the coronary ostia.

RESULTS

This approach resulted in a 100% overall technical success rate, excellent handling properties and precise positioning. The time taken to implant the 2 valved stents was 59 ± 12 min. Flow of the left and right coronary arteries did not significantly decrease after the stents were implanted (330.4 ± 12.06 ml/min vs 289.4 ± 5.29 ml/min, P < 0.001; 376.8 ± 10.5 ml/min vs 350.0 ± 14.5 ml/min; P < 0.001). We found no obvious regurgitation and perivalvular leakage; nor did the gradients of the aortic and mitral valves as well as of the left ventricular outflow tract increase significantly. The final angiographic examination and profile of the coronary opening confirmed the good position of the valved stents, the exclusion of the aneurysm and the patency of both coronary arteries.

CONCLUSIONS

These findings indicate the potential for combined transcatheter aortic root and mitral valve replacement in treating aortic root pathologies. In future, in vivo studies are expected to validate this approach and ascertain its durability.

Для цитирования: Zhao Yuxi, Li Tao, Wu Mingwei, Zeng Zhaoxiang, Gao Minxin, Bao Xianhao, Xu Ziyi, Zhou Jian, Feng Jiaxuan, Feng Rui, Jing Zaiping. Simultaneous transcatheter treatment of ascending aortic aneurysm with aortic and mitral regurgitation: an in vitro study.. Interactive cardiovascular and thoracic surgery. 2021. DOI: 10.1093/icvts/ivab101.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.