DOI 10.1080/17434440.2026.2710388
Percutaneous left ventricular assist device in cardiogenic shock associated with and without acute myocardial infarction: a real-world retrospective cohort study.
Аннотация
BACKGROUND
Percutaneous left ventricular assist devices (pLVAD, such as Impella), are increasingly used for cardiogenic shock (CS). Outcomes may differ between acute myocardial infarction-related CS (AMI-CS) and non-AMI CS due to differing pathophysiology and trajectories.
METHODS
Using the USA TriNetX Network (2016-2024), we identified adults with CS treated withpLVAD. AMI-CS was defined by MI within seven days of implantation; non-AMI CS included all patients with CS not attributable to acute MI, representing heterogeneous etiologies such as decompensated cardiomyopathy, myocarditis, valvular failure, pulmonary vascular causes, and arrhythmic shock. Patients with recent coronary artery bypass graft (CABG) were excluded. Propensity matching produced two balanced cohorts (n = 2,026 each).
RESULTS
Among 6,873 AMI-CS and 4,521 non-AMI CS patients, matched groups were similar (mean age 63 years, 26% female). AMI-CS had higher mortality at 30 days (hazard ratio [HR] 1.19, p = 0.002), 90 days (HR 1.13, p = 0.02), and 180 days (HR 1.14, p = 0.007). Heart failure (HF) exacerbations (HR 1.21, p < 0.001) and pulmonary edema (HR 1.23, p = 0.005) were also more common in AMI-CS. Stroke, ventricular arrhythmias, cardiac arrest, acute kidney injury, major bleeding, vascular complications, and hemodialysis were comparable.
CONCLUSION
AMI-CS patients supported with pLVAD experienced higher mortality and greater HF-related morbidity than non-AMI CS.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.