Bubble Trouble: The Thermal Risk of Ureteral Laser Lithotripsy in an Air Bubble.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Martin Gabriel E, Richards Kallan, Jhang Daniel, Battikha Adel, Betdashtoo Ninous, Park Jongwan, Ajayi Faith, Okhunov Zhamshid, Baldwin D Duane
Аннотация
INTRODUCTION: During ureteroscopic laser lithotripsy (URSL), unsafe elevations in ureteral temperature are thought to contribute to ureteral stricture formation. Cellular damage begins at 43°C, and proteins denature instantly at 60°C. Air bubbles may inadvertently migrate into the collecting system during URSL. The purpose of this study was to evaluate ureteral temperatures during simulated URSL in an air bubble with a holmium (Ho:YAG) and thulium fiber laser (TFL).
METHODS: BegoStone phantoms were positioned distal to the ureteropelvic junction in a model kidney and ureter submerged in a 35°C saline bath. A 7.5 Fr flexible ureteroscope was advanced to the stone, and continuous saline irrigation at 22°C was maintained at 15 mL/min. A needle thermocouple positioned 2 mm from the laser fiber tip monitored ureteral wall temperature. A 200 µm laser fiber was then advanced until it contacted the stone. The Ho:YAG and TFL were used in 5 trials each to ablate stones at 3, 10, and 20 W in saline and in a 3 mL air bubble.
RESULTS: In the control condition, neither laser exceeded 43°C at 3, 10, or 20 W. In an air bubble, ureteral temperatures exceeded 43°C after 1 second at 3 W for both TFL and Ho:YAG. At 10 W, the Ho:YAG exceeded 43°C after 1 second, while the TFL exceeded 60°C. The average temperature exceeded 60°C at 20 W for both lasers after 1 second. Both lasers generated much higher temperatures in an air bubble than in saline at all time points (p = 0.008), with maximum temperature exceeding 250°C for both lasers.
CONCLUSION: In this benchtop model, laser activation in an air bubble for 1 second exceeded the thermal safety threshold at ≥ 3 W and risks protein denaturation at ≥ 20 W. This study highlights the importance of avoiding air bubbles during URSL to maintain safe ureteral temperatures. Further studies are needed to validate these findings in vivo.
METHODS: BegoStone phantoms were positioned distal to the ureteropelvic junction in a model kidney and ureter submerged in a 35°C saline bath. A 7.5 Fr flexible ureteroscope was advanced to the stone, and continuous saline irrigation at 22°C was maintained at 15 mL/min. A needle thermocouple positioned 2 mm from the laser fiber tip monitored ureteral wall temperature. A 200 µm laser fiber was then advanced until it contacted the stone. The Ho:YAG and TFL were used in 5 trials each to ablate stones at 3, 10, and 20 W in saline and in a 3 mL air bubble.
RESULTS: In the control condition, neither laser exceeded 43°C at 3, 10, or 20 W. In an air bubble, ureteral temperatures exceeded 43°C after 1 second at 3 W for both TFL and Ho:YAG. At 10 W, the Ho:YAG exceeded 43°C after 1 second, while the TFL exceeded 60°C. The average temperature exceeded 60°C at 20 W for both lasers after 1 second. Both lasers generated much higher temperatures in an air bubble than in saline at all time points (p = 0.008), with maximum temperature exceeding 250°C for both lasers.
CONCLUSION: In this benchtop model, laser activation in an air bubble for 1 second exceeded the thermal safety threshold at ≥ 3 W and risks protein denaturation at ≥ 20 W. This study highlights the importance of avoiding air bubbles during URSL to maintain safe ureteral temperatures. Further studies are needed to validate these findings in vivo.