Percutaneous interventions on the hemodialysis reliable outflow vascular access device.
Аннотация
PURPOSE: To determine the outcomes of percutaneous interventions for prolonging the patency of the Hemodialysis Reliable Outflow (HeRO) device.
MATERIALS AND
METHODS
Between January 2007 and August 2011, 73 percutaneous interventions were performed on 26 HeRO devices in 25 patients. The graft was implanted in the upper arm with the outflow catheter tip in the superior vena cava or right atrium. Procedural reports, angiographic images, and clinical notes were retrospectively reviewed. The primary and secondary patency rates after intervention were calculated using the Kaplan-Meier method.
RESULTS
The mean time from HeRO implantation to initial dysfunction or thrombosis was 171 days. In 60 (82%) procedures, the HeRO device was thrombosed. An intragraft stenosis was the most common lesion identified (59%; n = 43) followed by an arterial anastomosis stenosis identified in 18% (n = 13). In 22% (n = 16) of procedures in which the HeRO device was thrombosed, an underlying cause was not identified after thrombectomy. The 3-, 6-, and 12-month primary patency rates after intervention were 47%, 37%, and 26% for first-time interventions. The secondary patency rates were 80%, 70%, and 64%. The only complication was pulmonary embolism resulting in death 2 days after HeRO thrombectomy.
CONCLUSIONS
Percutaneous interventions on thrombosed and failing HeRO devices yielded acceptable primary and secondary patency rates after intervention in these patients with few, if any, alternatives for hemodialysis access.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.