Thorax · 2026
Tool in lesion verification of shape-sensing robotic-assisted bronchoscopy with cone beam CT in sampling peripheral pulmonary nodules.
СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Chan Ley Taing, Lau Kelvin Kar Wing, Orton Christopher Michael, Temov Kire, Tana Anand, Baboolal Ilecia, Karir Ashish, Agaoglu Elif, Garner Justin, Kalyal Aisha, Lapuente Maria, Ttofia Foteini, Shah Pallav L
Аннотация
INTRODUCTION: Our aim was to evaluate the diagnostic and safety performance of shape-sensing robotic-assisted bronchoscopy (ssRAB) with cone beam CT (CBCT) for the biopsy of pulmonary nodules. Additional analysis was performed to assess outcomes for small nodules and those close to the fissure, pleura or mediastinum.
METHODS: This single arm, multicentre prospective study enrolled 200 subjects with suspicious pulmonary nodules. Each subject underwent a ssRAB procedure with CBCT and was subsequently followed up. The primary outcome was tool-in-lesion (TIL) confirmed with CBCT. Further endpoints included diagnostic outcomes and rate of adverse events.
RESULTS: Of 200 subjects recruited, 198 subjects had a successful biopsy (whereby lesion was reached and sample was taken) and 97% completed the required follow-up. The median size of the nodules was 13 mm; 26.8% (60/224) have a positive bronchus sign and 181 (80.8%) were located in the outer two-thirds of the lung. TIL was obtained in 99.0% (198/200). The strict diagnostic yield was 85% (170/200) with diagnostic accuracy of 92.0% (184/200) and sensitivity for malignancy of 95.5% (147/154). Diagnostic accuracy for nodules under 20 mm size, within 5 mm from a critical structure (eg, heart, aorta or main pulmonary artery) or from the pleura was 88.2% (172/195), 100% (11/11) and 93.3% (56/60), respectively. There were four (2%) serious procedure-related adverse events, with one patient (0.5%) suffering a pneumothorax.
CONCLUSION: ssRAB with CBCT can effectively reach and biopsy small pulmonary nodules, including perifissural, peripleural and paramediastinal lesions with a strong safety profile.
TRIAL REGISTRATION NUMBER: NCT05867953.
METHODS: This single arm, multicentre prospective study enrolled 200 subjects with suspicious pulmonary nodules. Each subject underwent a ssRAB procedure with CBCT and was subsequently followed up. The primary outcome was tool-in-lesion (TIL) confirmed with CBCT. Further endpoints included diagnostic outcomes and rate of adverse events.
RESULTS: Of 200 subjects recruited, 198 subjects had a successful biopsy (whereby lesion was reached and sample was taken) and 97% completed the required follow-up. The median size of the nodules was 13 mm; 26.8% (60/224) have a positive bronchus sign and 181 (80.8%) were located in the outer two-thirds of the lung. TIL was obtained in 99.0% (198/200). The strict diagnostic yield was 85% (170/200) with diagnostic accuracy of 92.0% (184/200) and sensitivity for malignancy of 95.5% (147/154). Diagnostic accuracy for nodules under 20 mm size, within 5 mm from a critical structure (eg, heart, aorta or main pulmonary artery) or from the pleura was 88.2% (172/195), 100% (11/11) and 93.3% (56/60), respectively. There were four (2%) serious procedure-related adverse events, with one patient (0.5%) suffering a pneumothorax.