Optimizing radiofrequency-assisted lead penetration into swine interventricular septum using pacing leads: A systematic in vitro assessment.
Аннотация
BACKGROUND
Left bundle branch area pacing is a promising physiological pacing modality, but fibrotic or calcified interventricular septal tissue can prevent lead advancement. Radiofrequency (RF) energy has been applied in transseptal puncture and lead extraction, but its optimal use for pacing lead penetration has not been systematically evaluated.
OBJECTIVE
We aimed to assess the safety and efficacy of different RF energy settings for facilitating pacing lead penetration through the interventricular septum in an ex vivo swine model.
METHODS
Freshly excised swine hearts (n = 6) were immersed in a temperature-controlled (37°C) circulating saline bath. The SelectSecure 3830 lead was positioned manually against the septum under fluoroscopic guidance. RF energy was delivered using an ERBE VIO300S electrosurgical generator in cutting mode across 12 combinations: 4 energy levels (10, 20, 30, and 55 W) and 3 durations (1, 2, 3 s). Additionally, spark mode (20, 30 W) was tested with 3 times and 5 times of sparking. Each combination was tested 6 times. Outcome metrics included penetration success (rupture of the right ventricular [RV] septal endocardium), safety markers (significant bubble formation, tissue charring), and histological assessment of myocardial injury.
RESULTS
At 10 W, penetration was not achieved in any trial. At 20 W, application for 1-2 seconds achieved an 88.9% success rate with minimal collateral injury. In contrast, power ≥30 W or durations >2 seconds significantly increased bubble formation (up to 94%) and deep myocardial necrosis. Histology confirmed that 20 W/1-2 seconds produced sharp cleavage of fibrotic tissue with preservation of deeper myocardial architecture. Spark mode was less stable and reproducible compared with optimized continuous settings.
CONCLUSION
RF-assisted pacing lead delivery using 20 W for 1∼2s represents an effective and safe strategy for overcoming superficial septal fibrosis in ex vivo settings. This technique may offer a viable rescue option during challenging left bundle branch area pacing procedures.