DOI 10.3390/jcm12196381

Медицинская библиотека · Journal of clinical medicine · 2023

Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders.

Espinoza Gustavo, Iglesias Katheriene, Parra Juan C, Rodriguez-Una Ignacio, Serrano-Gomez Sergio, Prada Angelica M, Galvis Virgilio

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study evaluated the agreement and reproducibility of ACA measurements obtained using the built-in software of the CASIA2 (Version 3G.1) and the measurements derived from expert clinicians.

METHODS

Healthy volunteers underwent ophthalmological evaluation and AS-OCT examination. ACA measurements derived from automated and manual SS location were obtained using the CASIA2 automated software and clinician identification, respectively. The intraobserver, interobserver reproducibility, CASIA2-human grader reproducibility and CASIA2 repeatability were assessed using intraclass correlation coefficients (ICCs).

RESULTS

The study examined 58 eyes of 30 participants. The CASIA2 software showed excellent repeatability for all ACA parameters (ICC > 0.84). Intraobserver, interobserver, and CASIA2-human grader reproducibility were also excellent (ICC > 0.87). Interobserver agreement was high, except for nasal TISA500, differing between observers 1 and 2 (p < 0.05). The agreement between CASIA2 measurements and human graders was high, except for nasal TISA500, where observer 1 values were smaller (p < 0.05).

CONCLUSION

The CASIA2 built-in software reliably measures ACA parameters in healthy individuals, demonstrating high consistency. Although a small difference was observed in nasal TISA500 measurements, interobserver and CASIA2-human grader reproducibility remained excellent. Automated SS detection has the potential to facilitate evaluation and monitoring of primary angle closure disease.

Для цитирования: Espinoza Gustavo, Iglesias Katheriene, Parra Juan C, Rodriguez-Una Ignacio, Serrano-Gomez Sergio, Prada Angelica M, Galvis Virgilio. Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders.. Journal of clinical medicine. 2023. DOI: 10.3390/jcm12196381.