The diagnostic accuracy of AI-assisted diabetic retinopathy screening in primary care: a prospective validation study.

Krogh Malene, Ørskov Marie, Nørgaard Thomas Lohne, Germund Nielsen Marie, Jensen Martin Bach, Jensen Morten Sig Ager, Vorum Henrik, Kristensen Jette Kolding

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

This study investigated the diagnostic accuracy of AI-assisted diabetic retinopathy screening in primary care, using ophthalmologist-led screening as the reference standard.

METHODS

Patients with type 2 diabetes attending routine appointments at 10 primary care clinics underwent AI-assisted screening, followed by re-screening at an ophthalmology clinic. The quality of fundus images captured in primary care was independently assessed, and diagnostic accuracy was evaluated by comparing AI-assisted results with ophthalmologist results, including sensitivity, specificity, PPV, NPV, and AUC. Two analyses were conducted: one including all images and one excluding those of poor quality.

RESULTS

Among 183 patients (336 images), 18.6% of images were classified as poor quality. When all images were included, the AI-assisted screening achieved a sensitivity of 73.7%, specificity of 90.2%, PPV of 31.1%, NPV of 98.3%, and AUC of 0.82. Excluding poor-quality images improved sensitivity to 80.0%, NPV to 98.7%, and AUC to 0.84. Additional ocular findings unrelated to diabetic retinopathy were observed in 96 patients, including confirmed or non-specific signs of glaucoma, cataract, age-macular degeneration, benign nevus and reduced visual acuity.

CONCLUSION

AI-assisted screening in primary care shows potential for clinical application, but further validation in larger populations and improvements in image quality are needed before clinical implementation.

Для цитирования: Krogh Malene, Ørskov Marie, Nørgaard Thomas Lohne, Germund Nielsen Marie, Jensen Martin Bach, Jensen Morten Sig Ager, Vorum Henrik, Kristensen Jette Kolding. The diagnostic accuracy of AI-assisted diabetic retinopathy screening in primary care: a prospective validation study.. Scandinavian journal of primary health care. 2026. DOI: 10.1080/02813432.2026.2683982.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.