Documenta ophthalmologica. Advances in ophthalmology · 2026

Impact of pupil dilation on IOL power calculation based on biometric measurements with Heidelberg Anterion.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Marquina-Martín Sara, Segura-Calvo Francisco J, Puzo-Bayod Martín, Jiménez-García Marta, Castro-Alonso Francisco J, UFR-HUMS Group Zaragoza

Аннотация

PURPOSE: To evaluate the impact of pharmacologic pupil dilation (PD) on biometric measurements and intraocular lens (IOL) power calculation using different generation formulas in patients undergoing cataract surgery.

METHODS: This prospective study included 126 eyes from 75 patients. Biometric data were obtained before and after pharmacologic dilation using the Anterion Swept-Source OCT biometer (Heidelberg Engineering). Parameters recorded included axial length (AL), mean keratometry (Km), anterior chamber depth (ACD), lens thickness (LT), and white-to-white (WTW) distance. IOL power was calculated using six formulas: SRK/T, Haigis, Barrett Universal II (BUII), Kane, Hoffer QST, and Okulix. Statistical comparisons of biometric values and predicted IOL power before and after dilation were conducted using mixed-effects models.

RESULTS: No significant changes were observed in AL, CCT, LT or Km after dilation. However, ACD, WTW distance and PD showed statistically significant increases (both p < 0.01). IOL power estimations based on SRK/T, Haigis, Hoffer QST, BUII and Kane were not significantly affected after dilation. In contrast, Okulix formulas demonstrated statistically significant variations in calculated IOL power post-dilation (p < 0.0001).

CONCLUSIONS: Pharmacologic pupil dilation induces changes in ACD and WTW measurements obtained using the Anterion Swept-Source biometer, thereby influencing IOL power estimation in formulas incorporating these parameters. Ray-tracing Okulix formula is more sensitive to dilation, whereas third-generation formulas remain more stable. Consistency in pupil status during biometry acquisition is essential to ensure refractive predictability.