[Explantation of the CyPass implant in a case series of patients with corneal decompensation].
Аннотация
BACKGROUND
In microinvasive glaucoma surgery (MIGS) the CyPass microstent (Alcon Inc.) was implanted in the suprachoroidal space to increase the uveoscleral outflow, thereby lowering the intraocular pressure. Due to the excessive loss of corneal endothelial cells, the implant was withdrawn from the global market in August 2018.
METHODS
A loss of corneal endothelial cells was observed in 6 patients (8 eyes in total) with early to moderate primary open-angle glaucoma 3 years after CyPass implantation. One eye developed a bullous keratopathy. In all patients an implant migration in the anterior chamber compared to the original postoperative position was detected, with more than one ring of the CyPass implant being visible in the anterior chamber.
RESULTS
In all eight cases the CyPass implant was removed. The local therapy with cortisone eye drops in combination with a 5% sodium chloride solution did not improve chronic corneal edema in one patient. Therefore, a Descemet membrane endothelial keratoplasty (DMEK) was performed with postoperative improvement of visual acuity.
CONCLUSION
In patients with more than one visible ring of the CyPass implant in the anterior chamber and subsequent localized loss of corneal endothelial cells, CyPass trimming or removal is recommended to avoid the progression of generalized corneal endothelial cell reduction and the onset of refractory, chronic corneal edema.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.