MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3109/13645706.2013.826675

Медицинская библиотека · Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy · 2013

The OTSC® Proctology clip system for anal fistula closure: first prospective clinical data.

Prosst Ruediger L, Ehni Wolfgang, Joos Andreas K

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Anorectal fistulas represent a troublesome condition for both patient and surgeon. The OTSC® Proctology clip system is a new device for transanal anorectal fistula closure.

MATERIAL AND

METHODS

The OTSC® Proctology system was evaluated in a prospective clinical study undertaken in two surgical proctological centers. Ten patients (three women, seven men; median age 55 years, range 25-73 years) with nine transsphincteric and one suprasphincteric fistulas were enrolled into the study.

RESULTS

The median operation time was 30 minutes (range 20-45 minutes). There were no intraoperative technical or surgical complications. Postoperatively, no patient reported intolerable discomfort, immoderate pain or foreign body sensation in the anal region. At follow-up examination six months after surgery, nine out of ten patients had no clinical signs or symptoms of their previous fistula and were considered as healed (success rate 90%). In one patient persistency of the fistula was noted due to spontaneous early clip detachment on the third postoperative day.

DISCUSSION: These first data demonstrate the efficacy of the OTSC® Proctology system in the treatment of anal fistulas. Currently, more patients were enrolled into this prospective study to allow further evaluation of this new device and to assess its future role in relation to established surgical strategies for anorectal fistulas.

Для цитирования: Prosst Ruediger L, Ehni Wolfgang, Joos Andreas K. The OTSC® Proctology clip system for anal fistula closure: first prospective clinical data.. Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy. 2013. DOI: 10.3109/13645706.2013.826675.

Предметные рубрики MeSH

Подготовка и перевод публикации