Feasibility of Direct Oral Anticoagulant Dipstick Point-of-Care Testing in Patients with Acute Stroke and Transient Ischaemic Attack.
Аннотация
INTRODUCTION
In patients with acute stroke, the plasma concentrations of direct oral anticoagulants (DOACs) need to be known for further treatment decisions. The point-of-care test DOAC Dipstick contains two test pads that determine levels of direct factor Xa (FXA) inhibitors and thrombin (THR) inhibitors in urine samples; thereby determining DOAC plasma levels of >30 ng/mL within 10 min. This study aimed to test the feasibility of using the DOAC Dipstick to detect the presence of DOACs in patients with acute stroke and transient ischaemic attack (TIA) during the early phase of intrahospital stroke management.
METHODS
Urine samples from patients with acute stroke and TIA with or without DOAC intake were obtained within 8 h after hospitalisation. The urine samples were tested using the DOAC Dipstick. After incubation, colour change on the DOAC Dipstick pads were analysed visually and using a semi-automated reader. Primary endpoints were the success rate of evaluable pad results and time to results after urine sampling.
RESULTS
From January 2023 to April 2024, 70 patients were recruited, 66 of whom fulfilled the inclusion criteria (39% female, mean age 73 ± 11.9 years). The median (interquartile range) times from urine sample collection to visual and reader test pad results were 20 (15-26) min and 24 (16-37) min, respectively. The test procedure was uncomplicated in most patients (n = 60). Complications in the test procedure were delays in urine sampling (n = 3), need for repeated testing due to dry strip (n = 2), no visual evaluability of the test result (n = 1), and abnormal urine colour (n = 1). The urine samples of all patients with prior DOAC intake (n = 26) were positive for DOACs while those for all patients without prior DOAC intake (n = 39) were negative for DOACs (1 patient was excluded because the urine sample was not collected in time).
CONCLUSION
The DOAC Dipstick rapidly and reliably detects DOACs in urine samples from patients with acute stroke and TIA.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.