DOI 10.3389/fcvm.2026.1790627

Медицинская библиотека · Frontiers in cardiovascular medicine · 2026

A systematic review and meta-analysis of axillary intra-aortic balloon pump as a bridge to advanced heart failure therapy.

Allaham Haytham, Kwiatkowski Sara, Alswayed Khalid, Bennet Samuel, Cong Olivia, Magder Laurence, Ramani Gautam, Chahal Diljon, Srivastava Mukta, Aalaei-Andabili Seyed-Hossein, Soares Cullen, Gupta Anuj

СтатьяОбзорПеревод в очереди

Аннотация

BACKGROUND

The axillary approach for intra-aortic balloon pump (IABP) placement offers an alternative to traditional transfemoral insertion, with the potential advantage of preserving mobility during temporary mechanical circulatory support.

METHODS

We performed a systematic review and meta-analysis of four retrospective studies evaluating outcomes in patients who received axillary IABP support.

RESULTS

Most patients supported with an axillary IABP were able to ambulate, with a pooled probability of 0.928 (95% CI: 0.811-0.999). The incidences of vascular complications (0.059; 95% CI: 0.001-0.137), stroke (0.022; 95% CI: 0.001-0.044), infection (0.037; 95% CI: 0.001-0.106), and bleeding (0.028; 95% CI: 0.005-0.052) were low. The most frequent adverse event was device failure (including kinking, rupture, migration, or malposition), with a pooled incidence of 0.314 (95% CI: 0.224-0.404).

CONCLUSION

Axillary IABP support is a feasible bridge to advanced heart failure therapies, enabling ambulation in most patients and facilitating transition to definitive treatment. While the overall safety profile is favorable, the relatively high rate of device failure represents a key limitation of this approach.

Для цитирования: Allaham Haytham, Kwiatkowski Sara, Alswayed Khalid, Bennet Samuel, Cong Olivia, Magder Laurence, Ramani Gautam, Chahal Diljon, Srivastava Mukta, Aalaei-Andabili Seyed-Hossein, Soares Cullen, Gupta Anuj. A systematic review and meta-analysis of axillary intra-aortic balloon pump as a bridge to advanced heart failure therapy.. Frontiers in cardiovascular medicine. 2026. DOI: 10.3389/fcvm.2026.1790627.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.