[Comparison of Procedural Parameters and Efficiency between Angio FFR and Wire-based FFR for Functional Severity Assessment].
Аннотация
PURPOSE: Angio FFR has attracted attention as a method for assessing the functional severity of intermediate coronary artery stenosis. However, detailed comparisons with wire-based FFR in terms of radiation dose and procedural efficiency remain limited. This study aimed to compare examination-related parameters and procedural efficiency between the two methods and to evaluate the clinical utility of angio FFR.
METHODS
This retrospective clinical study included patients who underwent functional assessment of coronary artery stenosis using either angio FFR or wire-based FFR during cardiac catheterization. The evaluated parameters were reference-point air kerma at the patient entrance reference point (Ka,r), air kerma-area product (PKA), fluoroscopy time, contrast volume, number of images, and angiography room stay time. Continuous variables were compared using median values and interquartile ranges.
RESULTS
Compared with the wire-based FFR group, the angio FFR group showed significantly lower Ka,r, fluoroscopy time, contrast volume, and angiography room stay time (p<0.05). No significant differences were observed in PKA (p=0.95) or the number of images (p=0.27).
CONCLUSION
Angio FFR was associated with lower Ka,r and improved procedural efficiency compared with wire-based FFR, suggesting its potential utility in functional assessment of coronary artery stenosis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.