Baruteau Alban-Elouen, Lambert Virginie, Riou Jean-Yves, Angel Claude-Yves, Belli Emre, Petit Jérôme

Аннотация

BACKGROUND: Closure of tubular patent ductus arteriosus remains a challenging procedure. Anecdotal use of Amplatzer Vascular Plug IV for tubular ductus closure has been reported but feasibility and safety in a consecutive patients' series remain unknown.

METHODS: We performed a monocenter prospective study at the Marie Lannelongue Hospital in Paris, France. From 2009 to 2014, a total of 47 patients (39 infants, 3 children, and 5 adults) underwent ductus closure with the Plug IV.

RESULTS: Ductus morphology was a type E in 34 (72.3%) patients and a type C in 13 (27.7%) patients. Ductus closure occurred in 39 (83.0%) infants at a median age of seven months (range: 3-23 months) and a median weight of 6.9 kg (range: 4.1-17.0 kg). A past history of prematurity and very low birth weight was found in 33 (70.2%) of them. Twelve (25.5%) patients had pulmonary hypertension. Mean Plug IV diameter was 1.9 ± 0.1 mm larger than the mean maximal ductus diameter. Early complete closure of the ductus was obtained in all patients. Early migration of an undersized Plug IV occurred in one (2.1%) patient and was suitable for percutaneous device retrieval. After a mean follow-up of 3.4 ± 1.4 years, all patients are alive and asymptomatic, no late complication occurred.

CONCLUSION: Transcatheter closure of tubular ductus with the Amplatzer Vascular Plug IV can be safe and effective, with a 100% early occlusion rate. This device, suitable for a 4F sheath, is a new alternative for tubular ductus closure in low-body-weight infants.