Pachyderm-Shape Guiding Catheters to Simplify BASILICA Leaflet Traversal.

Lisko John C, Babaliaros Vasilis C, Lederman Robert J, Khan Jaffar M, Rogers Toby, Greenbaum Adam B

Сравнительное исследованиеСтатьяНаблюдательное исследованиеResearch Support, N.I.H., IntramuralПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To test custom-shape guiding catheter curves intended to simplify aortic valve leaflet laceration.

BACKGROUND

Bioprosthetic Aortic Scallop Intentional Laceration to prevent Iatrogenic Coronary Artery obstruction (BASILICA) is an adjunct to transcatheter aortic valve replacement. It can be technically demanding using popular coronary guiding catheters and catheter-in-guide coaxial systems. New elephant-trunk shaped Pachyderm Left (PAL1,PAL2,PAL3, Launcher, Medtronic) and Right (PJR4) guiding catheters match the geometric requirements to engage aortic cusp hinge points. We evaluate whether these catheters ease BASILICA compared with conventional-shape coronary guiding catheters used in the BASILICA IDE trial.

METHODS

This is a single-center, consecutive, retrospective observational cohort of patients who underwent BASILICA for risk of TAVR-induced coronary obstruction defined as virtual valve-to-coronary distance <4 mm, immediately upon commercial availability of the new Pachyderm shaped catheters. Clinical, procedural, and angiographic details were abstracted from medical records of their index procedure, and were compared to adjudicated findings in the BASILICA IDE trial.

RESULTS

Nine leaflets in 6 patients were traversed and lacerated using BASILICA and Pachyderm curve guiding catheters for traversal, including three solo left and three doppio left and right leaflets. Leaflet traversal was universally successful. Leaflet time-to-leaflet traversal was shorter using Pachyderm catheters compared with the BASILICA IDE trial (8.3 min (5.6-15) vs 45 min (20, 61), p = 0.016). There were no deaths, strokes, or vascular complications.

CONCLUSIONS

The new Pachyderm shaped guiding catheters significantly hastened the leaflet time-to-traversal during BASILICA TAVR. We recommend they be used instead of conventional coronary-shaped guiding catheters.

Для цитирования: Lisko John C, Babaliaros Vasilis C, Lederman Robert J, Khan Jaffar M, Rogers Toby, Greenbaum Adam B. Pachyderm-Shape Guiding Catheters to Simplify BASILICA Leaflet Traversal.. Cardiovascular revascularization medicine : including molecular interventions. 2019. DOI: 10.1016/j.carrev.2019.05.033.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.