Impact of scapular notching on glenoid fixation in reverse total shoulder arthroplasty: an in vitro and finite element study.

Zhang Min, Junaid Sarah, Gregory Thomas, Hansen Ulrich, Cheng Cheng-Kung

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The high incidence of scapular notching in reverse total shoulder arthroplasty (RTSA) has spurred several methods to minimize bone loss. However, up to 93% of RTSAs accompanying scapular notching have been reported to maintain good implant stability for over 10 years. This study was conducted to investigate the relationship between scapular notching and glenoid fixation in RTSA.

METHODS

Cadaveric testing was performed to measure the notch-induced variations in strain on the scapular surface and micromotion at the bone-prosthesis interface during arm abductions of 30°, 60°, and 90°. Finite element analysis was used to further study the bone and screw stresses as well as the bone-prosthesis micromotion in cases with a grade 4 notch during complicated arm motions.

RESULTS

The notch resulted in an apparent increase in inferior screw stress in the root of the screw cap and at the notch-screw conjunction. However, the maximum stress (172 MPa) along the screw after notching is still much less than the fatigue strength of the titanium screw (600 MPa) under cyclic loading. The bone-prosthesis micromotion results did not present significant notch-induced variations.

CONCLUSIONS

Scapular notching will lead to few impacts on the stability of an RTSA on the glenoid side. This finding may explain the long-term longevity of RTSA in cases of severe scapular notching. The relationship between scapular notching and weak regions along the inferior screw may explain why fractures of the inferior screw are sometimes reported in patients with RTSA clinically.

Для цитирования: Zhang Min, Junaid Sarah, Gregory Thomas, Hansen Ulrich, Cheng Cheng-Kung. Impact of scapular notching on glenoid fixation in reverse total shoulder arthroplasty: an in vitro and finite element study.. Journal of shoulder and elbow surgery. 2020. DOI: 10.1016/j.jse.2020.01.087.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.