The Beckman DxI 800 prolactin assay demonstrates superior specificity for monomeric prolactin.
Аннотация
BACKGROUND
Commercially available prolactin immunoassays detect macroprolactin to variable degrees. Best practice requires laboratories to assess the cross-reactivity of their prolactin assay with macroprolactin, and where appropriate, introduce a screen for the presence of macroprolactin. Our policy has been to reanalyse hyperprolactinaemic samples following polyethylene glycol (PEG) precipitation and to report the resultant value as the monomeric prolactin content of the sample. The goal of this study was to determine the need to continue PEG precipitation when prolactin measurements with the Wallac AutoDELFIA were replaced by the Beckman DxI 800.
METHODS
A total of 317 apparently hyperprolactinaemic samples were analysed for prolactin using the Beckman DxI 800 and results compared with those determined with the PEG screening technique on the Wallac AutoDELFIA. Any samples demonstrating a discordance of >25% were re-analysed using gel filtration chromatography (GFC) for a definitive result.
RESULTS
The results indicate the Beckman DxI overestimates the prolactin concentration in 1%-2% of hyperprolactinaemic samples. PEG precipitation prior to analysis with the Wallac AutoDELFIA resulted in a 1% false negative diagnosis of hyperprolactinaemia.
CONCLUSIONS
Prolactin results can be reported directly from the DxI. When results are discordant with the clinical presentation, prolactin should be measured using GFC.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.