MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0315727

Медицинская библиотека · PloS one · 2025

Net buffer load during regional citrate anticoagulated continuous renal replacement therapy.

Forsal Innas, Pouchoulin Dominique, Roos Viktoria, Echeverri Jorge, Broman Marcus Ewert

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Regionally anticoagulated continuous renal replacement therapy with citrate is the first choice for critically ill patients with acute kidney injury. If citrate that reaches the patient exceeds the metabolic capacity, metabolic alkalosis will follow. Bicarbonate from the treatment fluids will also reach the patient and add to the bicarbonate load. Net buffer load is a parameter calculated by the dialysis machine software from the treatment fluid contents, the fluid flows and the filter properties. Each time the treatment settings are altered the net buffer load will be re-calculated. This parameter is integrated as a static safety parameter in the Prismax version 3 software, and will guide the operator on a possible development of metabolic alkalosis with the current settings.

METHODS

Algorithms for estimating the net buffer load were developed. Hourly clinical data from electronic medical records from 60 patients' continuous renal replacement treatments at the Adult Intensive Care Unit, Skåne University Hospital, Lund, Sweden was used to simulate net buffer load. The relation between net buffer load and acid base status at steady state was evaluated. Net buffer load was also calculated from three previous studies for comparison to our present cohort.

RESULTS

The mean net buffer load was +0.09 ±0.04 mmol/h/kg in the present cohort, and comparable to historic cohorts from literature. We could not establish a correlation to steady state arterial bicarbonate. The pre blood pump citrate containing replacement fluid flow rate was affecting net buffer load most of all parts of the dialysis circuit, with a r2 Pearson correlation coefficient of 0.65 (p<0.001).

CONCLUSION

The net buffer load parameter can alert the operator on the expected impact of the dialysis circuit on patient's acid base status. It was possible to calculate realistic net buffer load values during 60 CRRT treatments.

Для цитирования: Forsal Innas, Pouchoulin Dominique, Roos Viktoria, Echeverri Jorge, Broman Marcus Ewert. Net buffer load during regional citrate anticoagulated continuous renal replacement therapy.. PloS one. 2025. DOI: 10.1371/journal.pone.0315727.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.