Paced QRS morphology and lead location in left bundle branch area pacing assessed by photon-counting cardiac computed tomography.

Goto Kentaro, Miyazaki Shinsuke, Negishi Miho, Honda Masaki, Tateishi Ryo, Kawamura Iwanari, Nishimura Takuro, Yamao Kazuya, Tao Susumu, Takigawa Masateru, Sasano Tetsuo

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The relationship between the intra-septal lead position in left bundle branch area pacing (LBBAP) and paced QRS morphology remains incompletely understood. We used ultra high-resolution photon-counting computed tomography (PC-CT) to evaluate lead-tip location.

OBJECTIVE

We retrospectively analyzed 20 patients who underwent contrast-enhanced PC-CT after LBBAP implantations (9 with left bundle branch pacing [LBBP], 9 with left ventricular septal pacing [LVSP], and 2 with deep septal pacing [DSP]).

METHODS

Distances from the lead-tip to the LV and right ventricular (RV) endocardium (lead-tip to LV/RV), lead-tip to the presumed His bundle, and interventricular septal thickness were assessed.

RESULTS

The lead-tip to LV was significantly shorter in the LBBP group than in the other groups (LBBP: 1.8 ± 1.5 mm vs LVSP: 4.1 ± 1.4 mm vs DSP: 6.4 ± 0.4 mm, P < .01), with no significant differences in other anatomical parameters. Receiver operating characteristic analysis identified 3.2 mm as the optimal cutoff for LBB capture (area under the curve, 0.91; sensitivity, 100%; specificity, 81.8%). An initial q wave in lead V6 was observed exclusively in patients with a lead-tip to LV ≤3.2 mm (54.5% vs 0%, P < .01). The V1 R-wave peak time (V1RWPT) showed a moderate positive correlation with the lead-tip to RV (r = 0.636, P < 0.01). Pre-procedural echocardiographic septal thickness significantly underestimated the septal thickness at the penetration site (10.0 ± 1.9 mm vs 12.5 ± 3.8 mm, P < .01).

CONCLUSION

A lead-tip to LV distance ≤3.2 mm predicted LBB capture. An initial q wave in lead V6 may serve as a marker of lead positioning. V1RWPT was partially influenced by the lead-tip to RV.

Для цитирования: Goto Kentaro, Miyazaki Shinsuke, Negishi Miho, Honda Masaki, Tateishi Ryo, Kawamura Iwanari, Nishimura Takuro, Yamao Kazuya, Tao Susumu, Takigawa Masateru, Sasano Tetsuo. Paced QRS morphology and lead location in left bundle branch area pacing assessed by photon-counting cardiac computed tomography.. Heart rhythm. 2026. DOI: 10.1016/j.hrthm.2026.03.1945.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.